50 MHz Calibrator Amplitude Adjustment

This adjustment applies to the following models:

This adjustment adjusts the internal 50 MHz calibrator signal to be equal amplitude to an external 50 MHz, -25.00 dBm signal. The reference signal is provided by a calibrated, traceable oscillator.

Click here for troubleshooting.

Required Test Equipment

Test Equipment

Recommended Model Number

50 MHz Calibrator

Z5602B
   Opt H51 for Type-N models
   Opt H35 for 3.5 mm models
   Opt H24 for 2.4 mm models

Connection Setups

In the following setups, cables are designated as solid lines and direct connections are designated as dashed lines.

50 MHz Calibrator Amplitude Adjustment Setup