All X-Series models
This adjustment adjusts the internal 50 MHz calibrator signal to be equal amplitude to an external 50 MHz, -25.00 dBm signal. The reference signal is provided by a calibrated, traceable oscillator.
Click here for troubleshooting.
Test Equipment |
Recommended Model Number |
---|---|
50 MHz Calibrator |
Z5602B |
|
In the following setups, cables are designated as solid lines and direct connections are designated as dashed lines. |