Third Order Intermodulation Distortion Troubleshooting
Related Adjustments
If TOI fails to meet specifications, suspect
the A13 Front End Assembly. Next suspect the A2 Analog IF Assembly.
TOI is tested using low pass filters in Lowband.
The low pass filters are used to reduce the second harmonic of the signal
source. If a failure occurs in Lowband, the filters may be suspect.
|