If TOI fails to meet specifications, suspect the Front End Assembly. Next suspect the Analog IF Assembly.
TOI is tested using low pass filters in Lowband. The low pass filters are used to reduce the second harmonic of the signal source. If a failure occurs in Lowband, the filters may be suspect.