The goal of this calibration is to default the two arrays to optimize the IQ performance. This calibration adjusts the following arrays:
BB: IQ GAIN POWER BOOST POWER
BB: IQ RF RMS TARGET
No additional equipment required.
|
All test equipment requires a 30 minute warmup period to ensure accurate performance. |
Preset all test equipment and the signal generator.