Amplitude modulation frequency response (internal) measures the variation of the actual AM modulation depth at a given CW frequency when using the internal modulation source.
Test Equipment |
Critical Specifications for This Test |
Recommended Model |
Alternate Models |
Spectrum Analyzer |
AM Depth Accuracy |
N9030B Opt 5xx1 |
N9030A Opt 5xx1 |
|
|
|
If this performance test fails, perform the following steps in order:
Check the equipment setup (see above). If the setup is incorrect, make the necessary corrections and rerun the performance test.
Refer to the troubleshooting section of the service guide. If you do not have a printed copy of the service guide (Option OBW), one is available either on the CD-ROM that came with your signal generator shipment or search for your signal generator model number on the Keysight website.
If necessary, obtain service from Keysight Technologies. Refer to Contacting Keysight Technologies.