Harmonic Spurious

(N5161A / N5162A / N5166B / N5171B / N5172B / N5181A / N5181B / N5182A / N5182B / N5182N)

Harmonics are signal impurities that occur at integer multiples of a fundamental signal. For example, the 2nd harmonic frequency is two times the fundamental frequency. In this test, a spectrum analyzer measures the power of a harmonic component relative to the power of the fundamental frequency component.

The spectrum analyzer and the signal generator are both tuned to the harmonic frequency and a reference amplitude measurement is made. Then, the signal generator is tuned to the fundamental frequency with the power level unchanged. The spectrum analyzer measures the resulting harmonic response relative to the reference amplitude.

Required Test Equipment

Test Equipment

Critical Specifications for This Test

Recommended Model

Alternate Model(s)

Spectrum Analyzer

None

N9030B Opt 5xx1

N9030A Opt 5xx1
E444xA2

  1. Select any option of PXA spectrum analyzer (503, 508, 513, 526, 543, 544, 550) that has a maximum frequency greater than or equal to the maximum frequency of the unit under test. The N9030A/B Option 550 frequency range is adequate for all DUT frequency configurations.
  2. Select any model of PSA spectrum analyzer (E4440A, E4443A, E4445A, E4446A, E4447A, E4448A) that has a maximum frequency greater than or equal to the maximum frequency of the unit under test. The E4448A frequency range is adequate for all DUT frequency configurations.

Connections and Setup Procedures

  • All test equipment requires a 30 minute warmup period to ensure accurate performance.
  • In the following test setup, cables are designated as solid lines and direct connections are designated as dashed lines.
  1. Connect all test equipment as shown.
  2. Connect GPIB cables to all GPIB-controlled test equipment.
  3. While performing this test, follow all instructions on the controller display.

 

Test Setup

In Case of Difficulty

If this performance test fails, perform the following steps in order:

  1. Check the equipment setup (see above). If the setup is incorrect, make the necessary corrections and rerun the performance test.

  2. Refer to the troubleshooting section of the service guide. If you do not have a printed copy of the service guide (Option OBW), one is available either on the CD-ROM that came with your signal generator shipment or search for your signal generator model number on the Keysight website.

  3. If necessary, obtain service from Keysight Technologies. Refer to Contacting Keysight Technologies.