Internal Pulse Modulation Level Accuracy

M9484C with option PMR or PME

This test measures the relative amplitude of the RF Output when pulse modulated compared to the amplitude in CW mode.

The signal generator is first set to CW mode and its RF Output amplitude is measured on an oscilloscope to set the reference amplitude. The signal generator is then pulse modulated using its internal pulse generator and the oscilloscope measures the relative amplitude.

Required Test Equipment

Test Equipment

Recommended Model

Alternate Model

High performance oscilloscope

DSOZ634A1

  1. Can only be connected via LAN. Refer to the Connecting LAN Test Instruments section of the Getting Started topic.

Connection Setups

Measurement setup
For 1 and 2 channel M9484C

Uses Event 2 port.

Measurement setup
For 4-channel M9484C

Uses Event 1 port.