Calibration Constant Reset Utility

This utility resets all calibration constants stored in the FieldFox to their default values. Calibration constants contain the results for all adjustments except the Error Terms adjustment and the Station Calibration adjustment. The results of the Error Terms adjustment are stored separately from the calibration constants. The results of the Station Calibration adjustment are not stored in the FieldFox.

After this utility is executed, normally all adjustments required for the FieldFox model and options must be performed. For the N9912A and N9923A models, the Error Terms adjustment is not required following this utility unless hardware has been changed.

This utility may be used as a troubleshooting tool when it is suspected that existing calibration constant values may be causing operational problems.