This test measures the relative linearity error of the power sensor. All measurements are performed at 50 MHz. The reference power levels for linearity measurement are listed by UUT model series in the table below.
Click on the model link for a list of UUT models in that power sensor series.
UUT Model Series |
![]() N8480A |
![]() E9304A |
![]() 8480A |
8485D |
![]() N8480B |
![]() N8480H |
Power Level |
0 dBm |
0 dBm |
9 dBm |
–31 dBm |
30 dBm |
18 dBm |
This test uses a PSG source and/or attenuators to create power levels that are “deltas” from the reference level. Each delta is measured by the reference power sensor and the UUT. The power measurements made by the reference power sensor are limited to the nominal range around –14 dBm to –3 dBm where the reference power sensor is very linear and the measurement is very repeatable.
Instrument | Recommended product | Alternative product | N8480A, N8480B, 8480A, 8485D, E9304A | N8480H |
Signal Sources | ||||
Signal Generator | E8257D Opt 1EA or 1EU |
- | X | X |
Power Supply | N5745A | E3632A (Except N8480H) |
X | X |
Meter | ||||
Power Meter | N1914B | N1914A E4419B N1912A |
X | X |
Power Sensor[1] | ||||
Power Sensor | N8481A | 8481A | X For UUT N8481A (all except Opt 200), 8481A only |
|
Power Sensor | N8481A Opt 200 | 8481A Opt 001 | X For UUT N8481A Opt 200 only |
|
Power Sensor | N8482A | 8481A (For UUT N8480 Series) 8482A (For UUT 8480 Series) |
X For UUT N8482A, 8482A only |
|
Power Sensor | N8485A | 8485A | X For UUT N8485A, 8485A only |
|
Power Sensor | N8485A | N8481A, N8482A, N8487A, 8481A, 8482A, 8485A, 8487A |
X For UUT 8485D only |
|
Power Sensor | N8487A | 8487A | X For UUT 8487A only |
|
Power Sensor | N8481A | N8482A, N8485A, N8487A, 8481A, 8482A, 8485A, 8487A |
X For UUT E9304A Opt H18, E9304A,N8487A, N8488A, N8486AR, N8486AQ, N8480B only |
X |
Attenuators |
||||
3 dB High Power Fixed Attenuator | Aeroflex 40-3-43 |
- | X | |
10 dB High Power Fixed Attenuator | Aeroflex 40-10-43 | - | X | |
3 dB Fixed Attenuator |
8491A Opt 003 | 8491B Opt 003 |
X | |
6 dB Fixed Attenuator (3 required) |
8491A Opt 006 | 8491B Opt 006 | X | X (1 required) |
20 dB Fixed Attenuator | 8491A Opt 020 | 8491B Opt 020 | X | |
30 dB Attenuator | 11708A | - | X For UUT 8485D only |
|
10 dB Step Attenuator | 8496G | - | X | |
1 dB Step Attenuator | 8494G | - | X | |
Attenuator Driver | 11713B | 11713A | X | |
Miscellaneous devices | ||||
Directional Coupler | Pulsar Microwave Corporation C4-P4-411 |
- | X | |
Directional Bridge | 86205A |
- | X | |
Amplifier | Mini Circuit ZHL-100W-52 |
- | X | |
Amplifier | Mini Circuit ZHL-1-2W-N |
- | X | |
Filter | RLC Electronics BPF-750-50-10-5-N |
Alternative: Low Pass Filter 0955-0306 use with BNC (plug-m) to Type-N (f) 1250-1477 and BNC (jack-f) to Type-N (m) 1250-1476 |
X | X |
Adapters | ||||
3.5 mm (f) to Type-N (m) Adapter (2 required) | 1250-1744 |
- | X For UUT 8485A/8485D/N8485A only |
X |
3.5 mm (f) to Type-N (m) — used for connecting power sensors to the power meter reference port | 08485-60005 | - | X For UUT 8485A/8485D/N8485A only |
|
2.4 mm (f) to Type-N (m) — used for connecting power sensors to the power meter reference port | 08487-60001 | - | X For UUT N8487A/8487A/N8488A only |
|
2.4 mm (f) to Type-N (f) Adapter | 11903B | - | X | X |
Type-N (m) to Type-N (m) Adapter | 85032-60019 | - | X | |
APC-7 to Type-N (m) (2 required) | 85054-60032 | 11525A | X For UUT N8481A Opt 200 only |
|
Type-N (m) to Type-N (f) adapter | 33391C-K38 | - | X For UUT N8480B only |
|
Cables | ||||
Type-N (m to m) (3 required ) |
11500C | - | X | X |
Wire (from power supply to amplifier) | - | - | X | X |
[1] Supported power sensor options:
For UUT N8480A and 8480A, except N8487A, N8486AR, and N8486AQ:
For UUT 8485D:
For UUT E9304A, E9304A Opt H18, N8487A, N8488A, N8480B, and N8480H:
For UUT N8486AR and N8486AQ:
For UUT N8480A and 8480A, except N8487A, N8486AR, and N8486AQ:
For UUT 8485D, E9304A, E9304A Opt H18, N8487A, N8488A, N8480B, and N8480H:
For UUT N8486AR and N8486AQ:
For all UUT, except N8480H, 8485D, E9304A, and E9304A Opt H18:
For UUT 8485D, E9304A, and E9304A Opt H18:
For UUT N8480H:
Step 1:
Step 2: