Linearity

This test measures the relative linearity error of the power sensor. All measurements are performed at 50 MHz. The reference power levels for linearity measurement are listed by UUT model series in the table below.

Reference Power Levels

Click on the model link for a list of UUT models in that power sensor series.

UUT Model Series

8485D

Power Level

0 dBm

0 dBm

9 dBm

–31 dBm

30 dBm

18 dBm

This test uses a PSG source and/or attenuators to create power levels that are “deltas” from the reference level. Each delta is measured by the reference power sensor and the UUT. The power measurements made by the reference power sensor are limited to the nominal range around –14 dBm to –3 dBm where the reference power sensor is very linear and the measurement is very repeatable.

Required Test Equipment

Instrument Recommended product Alternative product N8480A, N8480B, 8480A, 8485D, E9304A N8480H
Signal Sources
Signal Generator E8257D Opt 1EA or 1EU
- X X
Power Supply N5745A E3632A
(Except N8480H)
X X
Meter
Power Meter N1914B N1914A
E4419B
N1912A
X X
Power Sensor[1]
Power Sensor N8481A 8481A X
For UUT N8481A (all except Opt 200), 8481A only
 
Power Sensor N8481A Opt 200 8481A Opt 001 X
For UUT N8481A Opt 200 only
 
Power Sensor N8482A 8481A (For UUT N8480 Series)
8482A (For UUT 8480 Series)
X
For UUT N8482A, 8482A only
 
Power Sensor N8485A 8485A X
For UUT N8485A, 8485A only
 
Power Sensor N8485A N8481A, N8482A,
N8487A, 8481A, 8482A, 8485A, 8487A
X
For UUT 8485D only
 
Power Sensor N8487A 8487A X
For UUT 8487A only
 
Power Sensor N8481A N8482A, N8485A,
N8487A, 8481A, 8482A, 8485A, 8487A
X
For UUT E9304A Opt H18, E9304A,N8487A, N8488A, N8486AR, N8486AQ, N8480B only
X
Attenuators
3 dB High Power Fixed Attenuator Aeroflex 40-3-43
-   X
10 dB High Power Fixed Attenuator Aeroflex 40-10-43 -   X
3 dB Fixed Attenuator
8491A Opt 003 8491B Opt 003
X  
6 dB Fixed Attenuator
(3 required)
8491A Opt 006 8491B Opt 006 X X
(1 required)
20 dB Fixed Attenuator 8491A Opt 020 8491B Opt 020   X
30 dB Attenuator 11708A - X
For UUT 8485D only
 
10 dB Step Attenuator 8496G - X  
1 dB Step Attenuator 8494G - X  
Attenuator Driver 11713B 11713A X  
Miscellaneous devices
Directional Coupler Pulsar Microwave Corporation
C4-P4-411
-   X
Directional Bridge 86205A
- X  
Amplifier Mini Circuit
ZHL-100W-52
-   X
Amplifier Mini Circuit
ZHL-1-2W-N
- X  
Filter RLC Electronics
BPF-750-50-10-5-N
Alternative:
Low Pass Filter 0955-0306 use with BNC (plug-m) to Type-N (f) 1250-1477 and BNC (jack-f) to Type-N (m) 1250-1476
X X
Adapters
3.5 mm (f) to Type-N (m) Adapter (2 required) 1250-1744
- X
For UUT 8485A/8485D/N8485A only
X
3.5 mm (f) to Type-N (m) — used for connecting power sensors to the power meter reference port 08485-60005 - X
For UUT 8485A/8485D/N8485A only
 
2.4 mm (f) to Type-N (m) — used for connecting power sensors to the power meter reference port 08487-60001 - X
For UUT N8487A/8487A/N8488A only
 
2.4 mm (f) to Type-N (f) Adapter 11903B - X X
Type-N (m) to Type-N (m) Adapter 85032-60019 - X  
APC-7 to Type-N (m) (2 required) 85054-60032 11525A X
For UUT N8481A Opt 200 only
 
Type-N (m) to Type-N (f) adapter 33391C-K38 - X
For UUT N8480B only
 
Cables
Type-N (m to m)
(3 required )
11500C - X X
Wire (from power supply to amplifier) - - X X

[1] Supported power sensor options:

Power Meter and Power Sensor Calibration (Channel A)

For UUT N8480A and 8480A, except N8487A, N8486AR, and N8486AQ:




For UUT 8485D:




For UUT E9304A, E9304A Opt H18, N8487A, N8488A, N8480B, and N8480H:




For UUT N8486AR and N8486AQ:




Power Meter and Power Sensor Calibration (Channel B)

For UUT N8480A and 8480A, except N8487A, N8486AR, and N8486AQ:




For UUT 8485D, E9304A, E9304A Opt H18, N8487A, N8488A, N8480B, and N8480H:




For UUT N8486AR and N8486AQ:




Linearity Test Setup

For all UUT, except N8480H, 8485D, E9304A, and E9304A Opt H18:




For UUT 8485D, E9304A, and E9304A Opt H18:




For UUT N8480H:

Step 1:




Step 2:




Standard Power Sensor Power Settling Timer




top