M9389A
This adjustment sweeps the frequency range and compares the measured power to the expected power. Using the comparison results, this procedure then makes adjustments to the absolute power accuracy of the ALC loop. This results in the very precise power accuracy of the instrument.
Test Equipment |
Recommended Models1 |
Power meter |
N1914A |
Power sensor |
E9304A H20 |
50 W terminator
|
909D |
Adapter, Type-N (f) to 3.5 mm (m) |
1250-1750 |
100 MHz reference |
M9300A |
Setup for Power Meter Calibration
Setup for Leveled Response Adjustment (M9380A, M9381A)
Setup for Leveled Response Adjustment (M9389A)