This utility resets the statistics (attenuator actuations, 50 MHz calibrator actuations, temperature ranges, etc.).
This statistical information is viewed on the CXA-m PXIe signal analyzer soft front panel when you press: System > More > Diagnostics > Show Hdwr Statistics.
The instrument on time is not allowed to be reset since this can have warranty and legal implications.
This utility should be run:
after replacing the attenuator
if a technician needs to reset the temperature extremes recorded by the instrument
The Statistics Reset Utility performs the following functions:
Resets mechanical attenuator actuations when the input attenuator is replaced.
Resets the 50 MHz switch actuations when the A9 Input Attenuator is replaced.
Resets the AC/DC switch actuations when the A9 Input Attenuator is replaced.
Resets the temperature extremes. A technician may wish to reset this counter when troubleshooting temperature related problems.