Amplitude Domain (Pulse)
Menu Path:
Determines the Amplitude Domain used for the Rise Time (Lower & Upper), Fall Time (Lower & Upper), and Pulse Width property percentage threshold levels.
- Voltage - All specified percentage threshold signal levels are calculated in voltage domain as a percentage of per pulse 100% Top Level in units of Volts.
- Power - All specified percentage threshold signal levels are calculated in power domain as a percentage of per pulse 100% Top Level in units of Watts.
Performance metrics affected include:
- Rise Time (s)
- Fall Time (s)
- Rise Edge (s)
- Fall Edge (s)
- Width (s)
- Off Time (s)
- PRI (s)
- PRF pseudo-random function (s)
- Duty Cycle (%)
- On Level (dBm deciBels referenced to a milliWatt: dB relative to 1 milliwatt dissipated in the nominal input impedance of the analyzer) (affected, but to a lesser extent)
- Peak Level (dBm) (affected, but to a lesser extent)
- Peak to Average (dB) (affected, but to a lesser extent)
Default selection: Voltage