(Read-Write) Set and query the method of computing the measured carrier power from the measured offset power in an Adjacent Channel Power measurement. |
|
Relevant Modes |
SA |
Parameters |
|
<bool> |
Choose from: 1or ON (Automatic) - dB or dBc value is computed by subtracting the measured carrier power from the measured offset power. 0 or OFF (Manual) - dB or dBc value is computed by subtracting the entered Ref Value from the measured offset power. Use [:SENSe]:ACPower:MREFerence to enter a value. |
Examples |
ACP:MREF:AUTO 1 |
Query Syntax |
[:SENSe]:ACPower:MREFerence:AUTO? |
Return Type |
Boolean |
Default |
ON |
Last Modified:
16-Aug-2012 |
New command |