[:SENSe]:AVERage:MODE <char>

(Read-Write) Set and query the averaging mode for NA measurements.

Relevant Modes

NA

Only Sweep averaging is allowed in CAT mode.

Parameters

 

<char>

Average Mode. Choose from:

SWEep - Each data point is based on the average of the same data point being measured over <n> consecutive sweeps. The average counter shows the number of previous sweeps that have been averaged together to form the current trace. When the counter reaches the specified count, then a ‘running average’ of the last <n> sweeps is displayed.

POINT - Each data point is measured <n> times and averaged before going to the next data point. On subsequent sweeps, averaging restarts by measuring each data point again <n> times. The average counter is not updated because data is not displayed until all the averages have been applied.

Examples

AVER:MODE POINT

Query Syntax

[:SENSe]:AVERage:MODE?

Return Type

Character

Default

SWEEP


Last Modified:

17-Jul-2012