(Read-Write) Set and query the averaging mode for NA measurements. |
|
Relevant Modes |
NA Only Sweep averaging is allowed in CAT mode. |
Parameters |
|
<char> |
Average Mode. Choose from: SWEep - Each data point is based on the average of the same data point being measured over <n> consecutive sweeps. The average counter shows the number of previous sweeps that have been averaged together to form the current trace. When the counter reaches the specified count, then a ‘running average’ of the last <n> sweeps is displayed. POINT - Each data point is measured <n> times and averaged before going to the next data point. On subsequent sweeps, averaging restarts by measuring each data point again <n> times. The average counter is not updated because data is not displayed until all the averages have been applied. |
Examples |
AVER:MODE POINT |
Query Syntax |
[:SENSe]:AVERage:MODE? |
Return Type |
Character |
Default |
SWEEP |
Last Modified:
17-Jul-2012 |
New command (6.00) |