(Read-Write) Set and query the correction coefficients (error terms) for the current calibration. |
|
Relevant Modes |
CAT, NA |
Parameters |
|
<char> |
Error term. Choose from: ES - Source match ER - Reflection tracking ED - Directivity All models EXCEPT N9912A: EL - Load Match ET - Transmission tracking EX - Isolation |
<resp1,stim2> |
Response port, Stimulus port. For N9912A, choose 1,1. For all other models, choose from 1 or 2 for each port.
|
<data> |
Two values per data point: (Real value, Imaginary value), separated by commas. |
Examples |
'Writes 3 data points of complex source match data CORR:COEF ES,1,1,2.626292E-01,1.823261E-01,2.537209E-01,1.943861E-01,2.448751E-01,2.059776E-01 |
Query Syntax |
[:SENSe]: CORRection:COEFficient[:DATA]? <char>, <resp1, stim2> |
Return Type |
Numeric |
Default |
Not Applicable |
Last Modified:
18-Oct-2012 |
Added new models |