[:SENSe]:CORRection:COEFficient[:DATA] <char>,<resp1,stim2>,<data>

(Read-Write) Set and query the correction coefficients (error terms) for the current calibration.

Relevant Modes

CAT, NA

Parameters

 

<char>

Error term. Choose from:

ES - Source match

ER - Reflection tracking

ED - Directivity

All models EXCEPT N9912A:

EL - Load Match

ET - Transmission tracking

EX  - Isolation

<resp1,stim2>

Response port, Stimulus port.

For N9912A, choose 1,1.

For all other models, choose from 1 or 2 for each port.

  • For ES, ER, or ED, the response port and the stimulus port must be the same.

  • For EL, ET, or EX, the response port and the stimulus port must be different.

<data>

Two values per data point: (Real value, Imaginary value), separated by commas.

Examples

'Writes 3 data points of complex source match data

CORR:COEF ES,1,1,2.626292E-01,1.823261E-01,2.537209E-01,1.943861E-01,2.448751E-01,2.059776E-01

Query Syntax

[:SENSe]: CORRection:COEFficient[:DATA]? <char>, <resp1, stim2>

Return Type

Numeric

Default

Not Applicable


Last Modified:

18-Oct-2012

Added new models