The effect of network analyzer error sources can be calculated as follows:
= N x S
where
is the error (or delta epsilon)
N is network analyzer uncertainty (in linear terms)
S is the sensitivity number (absolute value)
For example, assume PTFE (r=2.1) is measured at 10 GHz with a LOAD/AIR/SHORT calibration. At 10 GHz the fixed load/probe directivity contribution would be about 0.05. At 10 GHz for
r=2.1 the sensitivity is 8.1.
The measurement error due to the fixed load/probe directivity would be computed as:
= 0.05 x 8.1 = 0.4
Note that this is not total measurement error, only the portion due to the fixed load/probe directivity.