CALCulate:MEASure:SA Commands

Controls the marker settings used in the SA application.

CALCulate:MEASure:SA:MARKer:

BAND

       | FUNCtion

       | IBW

BDENsity

       | ACPR

           | [:STATe]

       | BW

       | DATA?

       | EQSPan

       | NOISe

           | [:STATe]

       | NPR

           | [:STATe]

       | POWer

           | BW

           | [:STATe]

       | TONE

           | BW

           | [:STATe]

           | TSPacing

BPOWer

       | DATA?

       | SPAN

       | [:STATe]

OCCBand

        | CENTer?

        | PERCent

        | POWer?

        | SPAN?

        | [:STATe]

SEARch

        | ACPR

        | NPR

Click on a keyword to view the command details.

See Also

Important: Learn about programming the reference marker.


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BAND:FUNCtion <enum>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB

(Read-Write) Set and read the SA marker function type.

Parameters

 

<cnum>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<enum>

Choose from:

MNOI - Marker noise.

BPOW - Band power.

BDEN - Band density.

OFF

Learn about these settings.

Examples

CALC:MEAS2:SA:MARK:BAND:FUNC MNOI

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BAND:FUNCtion?

Return Type

Enumeration

Default

OFF


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BAND:IBW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB

(Read-Write) Sets and reads the integration bandwidth marker. The IBW is used to determine total signal power within a specified frequency span. For example, to calculate the total power of a signal composed of 100 tones spaced 1 MHz apart over a 100 MHz span, the signal power would be integrated over an IBW of 100 MHz.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose an integration bandwidth.

Examples

CALC:MEAS2:SA:MARK:BAND:IBW 100e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BAND:IBW?

Return Type

Numeric

Default

Not Applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:ACPR[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB

(Read-Write) Sets and reads the ACPR density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band density marker OFF.

1 - ON - Turn band density marker ON.

Examples

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC2:MEAS:SA:MARK3:BDEN:ACPR:STAT 1

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:ACPR[:STATe]?

Return Type

Boolean

Default

0


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:BW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the bandwidth of the band density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a bandwidth.

Examples

CALC:MEAS2:SA:MARK:BDEN:BW 1e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:BW?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:DATA?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the band density level in dBm/Hz from the band density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Band density marker number. If unspecified, <n> is set to 1.

Examples

CALC:MEAS2:SA:MARK:BDEN:DATA?

calculate2:measure2:sa:marker2:bdensity:data?

Return Type

Numeric

Default

Not applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:EQSPan <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the frequency span used by Power Density to normalize the power.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a span.

Examples

CALC:MEAS:SA:MARK:BDEN:EQSPan 1e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:EQSPan?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:NOISe[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band density noise marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band density noise marker OFF.

1 - ON - Turn band density noise marker ON.

Examples

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC:MEAS:SA:MARK:BDEN:NOIS:STAT 1

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:NOISe?

Return Type

Boolean

Default

0


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:NPR[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB

(Read-Write) Sets and reads the NPR density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band density marker OFF.

1 - ON - Turn band density marker ON.

Examples

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC2:MEAS:SA:MARK3:BDEN:NPR:STAT 1

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:NPR[:STATe]?

Return Type

Boolean

Default

0


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:POWer[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band power density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band power density marker OFF.

1 - ON - Turn band power density marker ON.

Examples

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC:MEAS:SA:MARK:BDEN:POW:STAT 1

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:POWer?

Return Type

Boolean

Default

0


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:POWer:BW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the bandwidth of the band power density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a bandwidth.

Examples

CALC:MEAS:SA:MARK:BDEN:POW:BW 1e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:POWer:BW?

Return Type

Numeric

Default

1 MHz



CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE:BW <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the bandwidth of the band tone density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a bandwidth.

Examples

CALC:MEAS:SA:MARK:BDEN:TONE:BW 1e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE:BW?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band tone density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<bool>

Choose from:

0 - OFF - Turn band tone density marker OFF.

1 - ON - Turn band tone density marker ON.

Examples

'Create marker3 on that measurement

CALC2:MARK3 ON

'Make it a band density noise marker

CALC:MEAS:SA:MARK:BDEN:TONE:STAT 1

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE?

Return Type

Boolean

Default

0


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE:TSPacing <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the spacing of the band tone density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

<num>

Choose a spacing value.

Examples

CALC:MEAS:SA:MARK:BDEN:TONE:TSP 100e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE:TSPacing?

Return Type

Numeric

Default

100 MHz


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer:DATA?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the band power level from the band power marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Band power marker number.

Examples

CALC:MEAS2:SA:MARK:BPOW:DATA?

calculate2:measure2:sa:marker2:bpower:data?

Default

Not applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer:SPAN <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the frequency span of  the band power marker. This area is marked by two vertical dotted lines on the screen and the marker's y-axis value is set to the measured power value. Noise and power on the same marker share the same span.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

<num>

Choose a frequency span within the frequency range of the analyzer.

Examples

CALC:MEAS2:SA:MARK:BPOW:SPAN 1e6

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer:SPAN?

Return Type

Numeric

Default

1 MHz


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and reads the state of the band power marker. This command makes a band power marker from a generic marker. The generic marker must first be created using: CALC:MEAS:MARK:STATe

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

<bool>

Choose from:

0 - OFF - Turn band power marker OFF.

1 - ON - Turn band power marker ON.

Examples

'Create marker3 on the specified measurement

CALC2:MEAS2:MARK3 ON

'Make it a band power marker

CALC:MEAS2:SA:MARK:BPOW:STAT 1

Query Syntax

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer?

Return Type

Boolean

Default

0

CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:OCCBand:CENTer?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the occupied bandwidth center frequency.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

Examples

CALC:MEAS:SA:MARK:OCCB:CENT?

calculate2:measure:sa:marker2:occband:center?

Default

Not applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:OCCBand:PERCent <num>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and returns the percentage of the band power to search for.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

<num>

Percentage value.

Examples

CALC:MEAS:SA:MARK:OCCB:PERC 99

calculate2:measure:sa:marker2:occband:percent 99

Query Syntax

CALCulate<ch>:SA:MARKer<n>:OCCBand:PERCent?

Return Type

Numeric

Default

99.0


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:OCCBand:POWer?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the occupied bandwidth power.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

Examples

CALC:MEAS:SA:MARK:OCCB:POW?

calculate2:measure:sa:marker2:occband:power?

Default

Not applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:OCCBand:SPAN?

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-only) Returns the occupied bandwidth span.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

Examples

CALC:MEAS:SA:MARK:OCCB:SPAN?

calculate2:measure:sa:marker2:occband:span?

Default

Not applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:OCCBand[:STATe] <bool>

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Read-Write) Sets and returns the occupied bandwidth on/off state.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number.

<bool>

Choose from:

0 - OFF - Turns occupied bandwidth OFF.

1 - ON - Turns occupied bandwidth ON.

Examples

CALC:MEAS:SA:MARK:OCCB:STAT 1

calculate2:measure:sa:marker2:occband:state 1

Query Syntax

CALCulate<ch>:SA:MARKer<n>:OCCBand[:STATe]?

Return Type

Boolean

Default

0

Note: If occupied band state is turned ON, then Band Power or Band Noise is turned OFF.


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:SEARch:ACPR

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Write-only) Executes the search ACPR density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

Examples

CALC:MEAS:SA:MARK:SEAR:ACPR

calculate2:measure:sa:marker2:search:acpr

Query Syntax

Not Applicable

Default

Not Applicable


CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:SEARch:NPR

Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B)

(Write-only) Executes the search NPR density marker.

Parameters

 

<ch>

Channel number of the measurement (optional).

<mnum>

Measurement number for each measurement.

<n>

Marker number. If unspecified, <n> is set to 1.

Examples

CALC:MEAS:SA:MARK:SEAR:NPR

calculate2:measure:sa:marker2:search:npr

Query Syntax

Not Applicable

Default

Not Applicable