Controls the marker settings used in the SA application.
CALCulate:MEASure:SA:MARKer: BAND | FUNCtion | IBW BDENsity | ACPR | [:STATe] | BW | DATA? | EQSPan | NOISe | [:STATe] | NPR | [:STATe] | POWer | BW | [:STATe] | TONE | BW | [:STATe] | TSPacing BPOWer | DATA? | SPAN | [:STATe] OCCBand | CENTer? | PERCent | POWer? | SPAN? | [:STATe] SEARch | ACPR | NPR |
Click on a keyword to view the command details.
See Also
Important: Learn about programming the reference marker.
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB (Read-Write) Set and read the SA marker function type. |
|
Parameters |
|
<cnum> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<enum> |
Choose from: MNOI - Marker noise. BPOW - Band power. BDEN - Band density. OFF |
Examples |
CALC:MEAS2:SA:MARK:BAND:FUNC MNOI |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BAND:FUNCtion? |
Return Type |
Enumeration |
Default |
OFF |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB (Read-Write) Sets and reads the integration bandwidth marker. The IBW is used to determine total signal power within a specified frequency span. For example, to calculate the total power of a signal composed of 100 tones spaced 1 MHz apart over a 100 MHz span, the signal power would be integrated over an IBW of 100 MHz. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<num> |
Choose an integration bandwidth. |
Examples |
CALC:MEAS2:SA:MARK:BAND:IBW 100e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BAND:IBW? |
Return Type |
Numeric |
Default |
Not Applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB (Read-Write) Sets and reads the ACPR density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<bool> |
Choose from: 0 - OFF - Turn band density marker OFF. 1 - ON - Turn band density marker ON. |
Examples |
'Create marker3 on that measurement CALC2:MARK3 ON 'Make it a band density noise marker CALC2:MEAS:SA:MARK3:BDEN:ACPR:STAT 1 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:ACPR[:STATe]? |
Return Type |
Boolean |
Default |
0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the bandwidth of the band density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<num> |
Choose a bandwidth. |
Examples |
CALC:MEAS2:SA:MARK:BDEN:BW 1e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:BW? |
Return Type |
Numeric |
Default |
1 MHz |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-only) Returns the band density level in dBm/Hz from the band density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Band density marker number. If unspecified, <n> is set to 1. |
Examples |
CALC:MEAS2:SA:MARK:BDEN:DATA? calculate2:measure2:sa:marker2:bdensity:data? |
Return Type |
Numeric |
Default |
Not applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the frequency span used by Power Density to normalize the power. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<num> |
Choose a span. |
Examples |
CALC:MEAS:SA:MARK:BDEN:EQSPan 1e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:EQSPan? |
Return Type |
Numeric |
Default |
1 MHz |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the state of the band density noise marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<bool> |
Choose from: 0 - OFF - Turn band density noise marker OFF. 1 - ON - Turn band density noise marker ON. |
Examples |
'Create marker3 on that measurement CALC2:MARK3 ON 'Make it a band density noise marker CALC:MEAS:SA:MARK:BDEN:NOIS:STAT 1 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:NOISe? |
Return Type |
Boolean |
Default |
0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) or N524xB models with Option S93070xB (Read-Write) Sets and reads the NPR density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<bool> |
Choose from: 0 - OFF - Turn band density marker OFF. 1 - ON - Turn band density marker ON. |
Examples |
'Create marker3 on that measurement CALC2:MARK3 ON 'Make it a band density noise marker CALC2:MEAS:SA:MARK3:BDEN:NPR:STAT 1 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:NPR[:STATe]? |
Return Type |
Boolean |
Default |
0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the state of the band power density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<bool> |
Choose from: 0 - OFF - Turn band power density marker OFF. 1 - ON - Turn band power density marker ON. |
Examples |
'Create marker3 on that measurement CALC2:MARK3 ON 'Make it a band density noise marker CALC:MEAS:SA:MARK:BDEN:POW:STAT 1 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:POWer? |
Return Type |
Boolean |
Default |
0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the bandwidth of the band power density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<num> |
Choose a bandwidth. |
Examples |
CALC:MEAS:SA:MARK:BDEN:POW:BW 1e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:POWer:BW? |
Return Type |
Numeric |
Default |
1 MHz |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the bandwidth of the band tone density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<num> |
Choose a bandwidth. |
Examples |
CALC:MEAS:SA:MARK:BDEN:TONE:BW 1e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE:BW? |
Return Type |
Numeric |
Default |
1 MHz |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the state of the band tone density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<bool> |
Choose from: 0 - OFF - Turn band tone density marker OFF. 1 - ON - Turn band tone density marker ON. |
Examples |
'Create marker3 on that measurement CALC2:MARK3 ON 'Make it a band density noise marker CALC:MEAS:SA:MARK:BDEN:TONE:STAT 1 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE? |
Return Type |
Boolean |
Default |
0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the spacing of the band tone density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
<num> |
Choose a spacing value. |
Examples |
CALC:MEAS:SA:MARK:BDEN:TONE:TSP 100e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BDENsity:TONE:TSPacing? |
Return Type |
Numeric |
Default |
100 MHz |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-only) Returns the band power level from the band power marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Band power marker number. |
Examples |
CALC:MEAS2:SA:MARK:BPOW:DATA? calculate2:measure2:sa:marker2:bpower:data? |
Default |
Not applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the frequency span of the band power marker. This area is marked by two vertical dotted lines on the screen and the marker's y-axis value is set to the measured power value. Noise and power on the same marker share the same span. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
<num> |
Choose a frequency span within the frequency range of the analyzer. |
Examples |
CALC:MEAS2:SA:MARK:BPOW:SPAN 1e6 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer:SPAN? |
Return Type |
Numeric |
Default |
1 MHz |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and reads the state of the band power marker. This command makes a band power marker from a generic marker. The generic marker must first be created using: CALC:MEAS:MARK:STATe |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
<bool> |
Choose from: 0 - OFF - Turn band power marker OFF. 1 - ON - Turn band power marker ON. |
Examples |
'Create marker3 on the specified measurement CALC2:MEAS2:MARK3 ON 'Make it a band power marker CALC:MEAS2:SA:MARK:BPOW:STAT 1 |
Query Syntax |
CALCulate<ch>:MEASure<mnum>:SA:MARKer<n>:BPOWer? |
Return Type |
Boolean |
Default |
0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-only) Returns the occupied bandwidth center frequency. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
Examples |
CALC:MEAS:SA:MARK:OCCB:CENT? calculate2:measure:sa:marker2:occband:center? |
Default |
Not applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and returns the percentage of the band power to search for. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
<num> |
Percentage value. |
Examples |
CALC:MEAS:SA:MARK:OCCB:PERC 99 calculate2:measure:sa:marker2:occband:percent 99 |
Query Syntax |
CALCulate<ch>:SA:MARKer<n>:OCCBand:PERCent? |
Return Type |
Numeric |
Default |
99.0 |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-only) Returns the occupied bandwidth power. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
Examples |
CALC:MEAS:SA:MARK:OCCB:POW? calculate2:measure:sa:marker2:occband:power? |
Default |
Not applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-only) Returns the occupied bandwidth span. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
Examples |
CALC:MEAS:SA:MARK:OCCB:SPAN? calculate2:measure:sa:marker2:occband:span? |
Default |
Not applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Read-Write) Sets and returns the occupied bandwidth on/off state. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. |
<bool> |
Choose from: 0 - OFF - Turns occupied bandwidth OFF. 1 - ON - Turns occupied bandwidth ON. |
Examples |
CALC:MEAS:SA:MARK:OCCB:STAT 1 calculate2:measure:sa:marker2:occband:state 1 |
Query Syntax |
CALCulate<ch>:SA:MARKer<n>:OCCBand[:STATe]? |
Return Type |
Boolean |
Default |
0 Note: If occupied band state is turned ON, then Band Power or Band Noise is turned OFF. |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Write-only) Executes the search ACPR density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
Examples |
CALC:MEAS:SA:MARK:SEAR:ACPR calculate2:measure:sa:marker2:search:acpr |
Query Syntax |
Not Applicable |
Default |
Not Applicable |
Applicable Models: All with Spectrum Analysis Options (S9x09xxA/B, S9x090A/B) (Write-only) Executes the search NPR density marker. |
|
Parameters |
|
<ch> |
Channel number of the measurement (optional). |
<mnum> |
Measurement number for each measurement. |
<n> |
Marker number. If unspecified, <n> is set to 1. |
Examples |
CALC:MEAS:SA:MARK:SEAR:NPR calculate2:measure:sa:marker2:search:npr |
Query Syntax |
Not Applicable |
Default |
Not Applicable |