SYSTem:CALibrate:NOISe Commands


Contains the settings to perform a Noise BW Calibration.

SYSTem:CALibrate:NOISe

BWCAL

     | DELete

     | CHANnel

     | BWIDth

     | DESCription

     | ACQuire

Click on a red keyword to view the command details.


SYSTem:CALibrate:NOISe:BWCAL:DELete

Applicable Models: NA520xA

(Command-Only) Removes the noise bandwidth XML file.

Parameters

Examples

SYST:CAL:NOIS:BWCAL:DEL

SYSTem:CALibrate:NOISe:BWCAL:CHANnel <chNum>

Applicable Models: NA520xA

(Write-Read) Determines port numbers to use for calibration. Determines appropriate stimulus settings.

Parameters

<chNum>

Channel number to be calibrated.

Examples

SYST:CAL:NOIS:BWCAL:CHAN 1

Query Syntax

SYSTem:CALibrate:NOISe:BWCAL:CHANnel?

Return Type

Numeric

Default

None

SYSTem:CALibrate:NOISe:BWCAL:BWIDth <bw1,bw2,bw3>

Applicable Models: NA520xA

(Write-Read) Specifies what bandwidths to calibrate.

Parameters

<string>

Cal Kit.

Examples

SYST:CAL:NOIS:BWCAL:BWID 3e6,4e6,5e6

Query Syntax

SYSTem:CALibrate:NOISe:BWCAL:BWIDth?

Return Type

Comma-separated list of doubles

Default

4e6

SYSTem:CALibrate:NOISe:BWCAL:DESCription?

Applicable Models: NA520xA

(Query-Only) Gives connection description (thru between ports X and Y with bandwidth limiting devices removed).

Parameters

Examples

SYST:CAL:NOIS:BWCAL:DESC?

Query Syntax

SYSTem:CALibrate:NOISe:BWCAL:DESCription?

Return Type

String

SYSTem:CALibrate:NOISe:BWCAL:ACQuire

Applicable Models: NA520xA

(Command-Only) Needs synchronous and asynchronous versions. Acquires the data and stores the results to the XML file..

Parameters

Examples

SYST:CAL:NOIS:BWCAL:ACQ