Message 338

Message

"338. {0} is outside of the calibrated temperature range. Measurements may not be accurate."

Description

If the temperature exceeds ±5° C from the temperature at which the oscilloscope frame was last adjusted/calibrated, this message cautions that the oscilloscope's measurements may not be accurate.

Either return the oscilloscope to an environment in the expected temperature range or re-adjust/calibrate the oscilloscope at the new temperature.