Default Conditions

Initial Settings, Settings that can be Saved/Recalled, Settings that can be Backed Up

The following table shows the following items.

√ : Settings that can be saved/recalled.

× : Settings that cannot be saved/recalled.

√ : Settings that can be backed up.

× : Settings that cannot be backed up.

SCPI: Settings that can be set by remote controller using a SCPI command.

Front: Settings that can be set by the front panel keys, the keyboard or the mouse.

Both: Settings that can be set by either of the above methods.

Setting Items

Default Value

Preset

*RST

Save/ Recall

Backup

Available method to set

Measurement parameter

Parameter 1

|Z|

×

Both

Parameter 2

θz

×

Both

Parameter 3

Rs

×

Both

Parameter 4

X

×

Both

List measurement/single-point measurement

Single-point measurement

×

Both

Measured point number on single-point measurement

1

×

Both

Active table number

1

×

Both

Measurement point setup table

Number of measurement points

1

×

Both

Oscillator

Frequency

1 [MHz]

×

Both

Level

-13 [dBm]

×

Both

Averaging factor

1

×

Both

Unit of oscillator level

dBm

×

Both

Rdc measurement

ON/OFF

OFF

×

Both

Offset cancel function

OFF

×

SCPI

Limit range

Lower limit value

0 [Ω]

×

Both

Upper limit value

10 [kΩ]

×

Both

Averaging factor

128

×

Both

Deviation measurement

ON/OFF

OFF

×

Both

Mode

DEV

×

Both

Reference value (center value)

0

×

Both

Delay time

Trigger delay time

0 [s]

×

Both

Measurement point delay time

0 [s]

×

Both

Display ON/OFF

ON

×

Both

Results display (measurement parameters 1 to 4, test signal level monitor, Rdc measurement)

All results display ON/OFF

ON

×

Both

Each result displays ON/OFF (except Rdc measurement result)

ON

×

Both

Number of digits

4

×

Both

Fixed decimal point mode ON/OFF

OFF

×

Both

Most significant digit

0

×

Both

Title display

ON/OFF

OfFF

×

Both

Title characters

" " (blank)

×

Both

Time and date display

ON/OFF

OFF

×

Both

Live/Stamp

Stamp

×

Both

Displayed items on the list measurement display

Item 1 (LABEL-1)

Measurement parameter 1

×

Both

Item 2 (LABEL-2)

Measurement parameter 2

×

Both

Item 3 (LABEL-3)

Measurement parameter 3

×

Both

Item 4 (LABEL-4)

Measurement parameter 4

×

Both

Trigger source

Internal

×

Both

Polarity of external trigger

Positive

×

Both

Trigger system continuous activation ON/OFF

ON

OFF

×

SCPI

Maximum data size for statistical analysis

3000

×

Both

Comparator (BIN sort) funtion ON/OFF

OFF

×

Both

Boundary between good BIN and no good BIN

10

×

Both

BIN1 to BIN13 setting

ON/OFF

OFF

×

Both

Sorting conditions 1 to 32

Measurement point number

1

×

Both

Parameter

(n=1 to 8)

Condition (n-1)*4+1

|Z|

×

Both

Condition (n-1)*4+2

θz

×

Both

Condition (n-1)*4+3

Rs

×

Both

Condition
(n-1)*4+4

X

×

Both

Limit mode (ABS/%/DEV)

ABS

×

Both

Limit

Nominal value

0

×

Both

Lower limit value

0

×

Both

Upper limit value

0

×

Both

BIN range mode (IN/OUT/ALL)

ALL

×

Both

BIN count function ON/OFF

Off

×

Both

Setting of 'Calibration Wizard'

Calibration

Low-Loss C

No Check

No effect

×

Front

DC Open

No check

No effect

×

Front

DC Short

No check

No effect

×

Front

DC Load

No check

No effect

×

Front

Rdc Limit

Rdc Limit

On (Check)

No effect

×

Front

Open

Min

1 [kΩ]

No effect

×

Front

Load

Max

55 [Ω]

No effect

×

Front

Min

45 [Ω]

No effect

×

Front

Short

Max

1 [Ω]

No effect

×

Front

Compensation

Open

Check

No effect

×

Front

Short

Check

No effect

×

Front

DC Open

No check

No effect

×

Front

DC Short

No check

No effect

×

Front

Rdc Limit

Rdc Limit

On (Check)

No effect

×

Front

Open

Min

1 [kΩ]

No effect

×

Front

Short

Max

1 [Ω]

No effect

×

Front

Test Fixture Selection

None

No effect

×

Front

Target table on measurement of data (All tables/Active table only)

Active table only

No effect

×

Front

Automatically saves the state into auto-recall file (autorec.sta) when the 'Calibration Wizard' is completed

No check

No effect

×

Front

Target table in measurement of data for calibration coefficient calculation (all tables/active tables only)

  • When data measurement is performed by SCPI command, the target table is always 'active table only' regardless of this setting.

Active table only (No check)

No effect

×

Front

Automatically saves the state into auto-recall file (autorec.sta) when the calibration coefficients are calculated.

  • When calculation of the calibration coefficients is performed by SCPI command, the state is not saved into auto-recall file (autorec.sta) regardless of this setting.

No check

No effect

×

Both

Calibration kit selection (16195B-7 mm standard/user defined)

16195B

×

Both

Custom Averaging (For both calibration and compensation)

No Check

Both

Minimum Count (For both calibration and compensation)

1

Both

User defined calibration kit

Impedance definition method (fixed/point-by-point)

Fixed

×

Both

Open

Equivalent parallel conductance (G)

0 [S]

×

Both

Equivalent parallel capacitance (Cp)

0 [F]

×

Both

Dc conductance

0 [S]

×

Both

Offset delay time

0 [s]

×

Both

Short

Equivalent series resistance (Rs)

0 [Ω]

×

Both

Equivalent series inductance (Ls)

0 [H]

×

Both

Dc resistance

0 [Ω]

×

Both

Offset delay time

0 [s]

×

Both

Load

Definition impedance parameter type

Rs and Ls

×

Both

Equivalent series resistance (Rs)

50 [Ω]

×

Both

Equivalent series inductance (Ls)

0 [H]

×

Both

Dc resistance

50 [Ω]

×

Both

Offset delay time

0 [s]

×

Both

Open compensation On/Off

OFF

×

Both

Short compensation On/Off

OFF

×

Both

Target table in measurement of data for compensation coefficient calculation (all tables/active table only)

  • When data measurement is performed by SCPI command, the target table is always 'active table only' regardless of this setting.

Active table only (No Check)

No effect

×

Front

Automatically saves the state into auto-recall file (autorec.sta) when the compensation coefficients are calculated.

  • When calculation of the calibration coefficients is performed by SCPI command, the state is not saved into auto-recall file (autorec.sta) regardless of this setting.

No check

No effect

×

Both

Selection of compensation kit definition values (predefined values/user defined values)

Predefined values (default)

×

Both

User defined values

Impedance definition method (fixed/point-by-point)

Fixed

×

Both

Open

Equivalent parallel conductance (G)

0 [S]

×

Both

Equivalent parallel capacitance (C)

0 [F]

×

Both

Dc conductance

0 [S]

×

Both

Short

Equivalent series resistance (Rs)

0 [Ω]

×

Both

Equivalent series inductance (Ls)

0 [H]

×

Both

Dc resistance

0 [Ω]

×

Both

Test fixture selection

None

×

Both

User fixture

Electrical length

0 [m]

×

Both

Name

"USER"

×

Both

Data transfer format

Data transfer format (ASCII/Binary)

ASCii

No effect

ASCii

×

×

SCPI

Byte order when data transfer format is set to binary (normal/swapped)

NORmal

No effect

NORmal

×

×

SCPI

Key lock

Front panel

OFF

×

×

Both

Keyboard

OFF

×

×

Both

Mouse

OFF

×

×

Both

Beeper

On/Off of beep to notify the completion of operation

ON

×

Both

Beep to notify an error/warning or sorting results

ON/OFF

ON

×

Both

Mode

FAIL

×

Both

Time and date of the internal clock

--------

   

×

Both

GPIB address

17

 

 

×

Front

Telnet Server

ON

 

 

×

Front

SICL LAN

ON

 

 

×

Front

Web Server

ON

 

 

×

Front

Service request enable register value

0

   

×

×

SCPI

Standard event status enable register value

0

   

×

×

SCPI

Operation status register

Enable register value

0

   

×

×

SCPI

 

Positive transition filter value

32767

   

×

×

SCPI

 

Negative transition filter value

0

   

×

×

SCPI

Questionable status enable register value

0

   

×

×

SCPI