:CALC:PAR{1-4}:EXPR

Syntax

:CALCulate:PARameter{1|2|3|4}[:MATH]:EXPRession[:STATe] {ON|OFF|1|0}

:CALCulate:PARameter{1|2|3|4}[:MATH]:EXPRession[:STATe]?

Description

Specifies whether to use deviation mode to display measurement results for each of the parameters 1 (:PAR1) through 4 (:PAR4). In deviation mode, measurement results are expressed as deviations relative to the reference value (specified with the :CALC:PAR{1-4}:EXPR:CENT command).

Variable

Parameter

State

Description

ON/OFF deviation mode.

Range

Select from the following:

  • ON or 1: Turns ON deviation mode

  • OFF or 0: Turns OFF deviation mode (i.e., instructs the instrument to display measurement results as absolute values)

Preset value

OFF or 0

Query response

{1|0}<newline><^END>

Related objects

:CALC:PAR{1-4}:EXPR:NAME

:CALC:PAR{1-4}:EXPR:CENT

Equivalent key

Display > Setup Prmtr Display > {Parameter 1|Parameter 2|Parameter 3|Parameter 4} > Math Mode > {ON|OFF}