:DATA:RCAD{1-3}?

Syntax

:DATA[:DATA]:RCAD{1|2|3}?

Description

Returns a set of measurement data (real number) for calculating the calibration coefficients to use during Rdc measurement. The instrument holds three sets of calculation data, which are obtained by measuring the OPEN, SHORT, and LOAD standards. To obtain a particular data set, suffix the command with the number that corresponds to the desired data set. The command suffix numbers associated with each data set are listed below. (Query only)

1: OPEN measurement data

2: SHORT measurement data

3: LOAD measurement data

Query response

{numeric}<newline><^END>

Equivalent key

No equivalent keys are available on the front panel.