:STATus:OPERation:PTRansition <numeric>
:STATus:OPERation:PTRansition?
Sets the value of the positive transition filter of operation status register.
Parameter |
<numeric> |
Description |
Value of positive transition filter |
Range |
0 to 32767 |
Preset value |
32767 |
Resolution |
1 |
If you specify a parameter value beyond the valid range, the command assumes a bit-by-bit logical product (AND) with 32767 (0x7fff).
{numeric}<newline><^END>
Trigger Mode > Ext Trig Polarity > Positive