:SENS<Ch>:CORR:CKIT:LOAD:PRE

Type

Command/Query

Syntax

:SENSe<Ch>:CORRection:CKIT:LOAD:PREal <Value>

:SENSe<Ch>:CORRection:CKIT:LOAD:PREal?

Description

This command sets/gets User Cal Standard value Dielectric Load Real for Permittivity measurement.

Variable

Parameter

Value

Description

User Cal Standard value Dielectric Load Real for Permittivity measurement

Data Type

Numeric type (Real)

Range

-1M ~ 1M

Preset Value

2.1

Unit

-

Resolution

-

Query Response

{numeric}<newline><^END>

Examples

Dim Var as Double
Var= -1000000
Ana.WriteString ":SENS1:CORR:CKIT:LOAD:PRE " & Str(Var), True
Ana.WriteString ":SENS1:CORR:CKIT:LOAD:PRE?", True
Var=Ana.ReadNumber

Related Commands

:SENS:CORR:CKIT:LOAD:OFFS

:SENS:CORR:CKIT:LOAD:PLF

:SENS:CORR:CKIT:LOAD:THIC

Equivalent Softkey

 Cal Compen > Calibration > Define Value > Load Real (er')

 Instrument Setup > Calibration > Define Value > Load Real (er')

COM Command Syntax (Internal Control Only)

SCPI.SENSE(Ch).CORRECTION.CKIT.LOAD.PREAL