Command/Query
:SENSe<Ch>:CORRection2:SHORt[:STATe] {ON|OFF|1|0}
:SENSe<Ch>:CORRection2:SHORt[:STATe]?
This command sets/gets the short fixture compensation On/Off status.
SENS:CORR1:SHOR causes an undefined header error.
(<type>: User Calibration=1, Temperature Compensation=3)
Parameter |
Selection Option |
Description |
Short fixture compensation On/Off status |
Data Type |
Boolean type (Boolean) |
Range |
ON|OFF|1|0 |
Preset Value |
OFF |
{1|0}<newline><^END>
Dim Var as Integer
Ana.WriteString ":SENS1:CORR2:SHOR ON", True
Ana.WriteString ":SENS1:CORR2:SHOR?", True
Var=Ana.ReadNumber
Cal > Fixture Compen > Short
SCPI.SENSE(Ch).CORRECTION(2).SHORT.STATE