:SENS<Ch>:CORR<Type>:ZME:OPEN

Type

Command/Query

Syntax

:SENSe<Ch>:CORRection<Type>:ZMEas:OPEN[:DATA] <Value>

:SENSe<Ch>:CORRection<Type>:ZMEas:OPEN[:DATA]?

Description

This command gets the actual impedance array of the open standard for the fixture compensation. <type>: User Calibration=1, Fixture Compensation=2, Temperature Compensation=3

 

The array size (number of point):

Variable

Parameter

Value

Description

Open standard impedance array

Data Type

Variant type Array (Range)

 

Query Response

{numeric}, ... , {numeric}<newline><^END>

Examples

See :SENS:CORR:ZME:LOAD

Related Commands

:SENS:CORR:ZME:SHOR

:SENS:CORR:ZME:LOAD

:SENS:CORR:ZME:OPEN:POIN

:SENS:CORR:ZME:POIN

Equivalent Softkey

No equivalent key is available on the front panel.

COM Command Syntax (Internal Control Only)

SCPI.SENSE(Ch).CORRECTION(2).ZMEAS.OPEN.DATA