Before starting the measurement, you must set the measurement parameters and sweep conditions according to your measurement requirements. This section describes the setup procedure for the following measurement.
Frequency characteristics of εr' - εr" - tanδ
First you should change the measurement conditions from the initial state of the E4991B as shown in the below table.
Parameter setting |
Setup example |
Initial state |
|
Measurement parameters |
Trace 1 |
εr' |
εr' |
Trace 2 |
εr" |
εr" |
|
Trace 3 |
tanδ |
tanδ |
|
Display formats |
Trace 1 |
Linear |
Linear |
Trace 2 |
Linear |
Linear |
|
Trace 3 |
Linear |
Linear |
|
Sweep parameter |
Frequency |
Frequency |
|
Sweep type |
Log |
Linear |
|
Source mode |
Voltage |
Voltage |
|
Oscillator level |
100 mV |
100 mV |
|
Sweep range (frequency) |
1 MHz to 1 GHz |
1 MHz to 3 GHz |
Press Display > Num of Traces > 3.
Press Meas > Trace 1 > εr'.
Press Format > Y-Axis > Linear.
Similarly, set Trace 2 > εr" with Y-Axis > Linear and Trace 3 > tanδ with Y-Axis > Linear.
Press Sweep Setup > Sweep Type > Log Freq.
Press Sweep Setup > Osc Level > Osc Unit > Voltage.
Press Sweep Setup > Osc Level > Voltage Level > 100 > k/m.
Press Start > Start > 1 > M/μ.
Press Stop > Stop > 1 > G/n.