Before starting the measurement, you must set measurement parameters and sweep conditions depending on your measurement requirements. This section describes the setup procedure for the following four measurements.
The basic procedure for setting the measurement conditions is illustrated below:
Change the measurement conditions from the initial state of the E4991B:
Setup example for this measurement:
Parameter setting |
Setup example |
Initial state |
|
Measurement parameters |
Trace 1 |
|Z| |
|Z| |
Trace 2 |
Ls |
θz |
|
Trace 3 |
Q |
Q |
|
Display formats |
Trace 1 |
Log |
Linear |
Trace 2 |
Linear |
Linear |
|
Trace 3 |
Linear |
Linear |
|
Measurement points |
201 points |
201 points |
|
Sweep parameter |
Frequency |
Frequency |
|
Sweep type |
Log |
Linear |
|
Source mode |
Current |
Voltage |
|
Oscillator level |
1 mA |
100 mV (2 mA) |
|
Sweep range (frequency) |
1 GHz to 3 GHz |
1 MHz to 3 GHz |
Preset the E4991B.
Press Display > Num of Traces > 3.
Press Meas > Trace 1 > |Z|.
Press Format > Y-Axis > Log.
Similarly, set Trace 2 > Ls with Y-Axis > Linear and Trace 3 > Q with Y-Axis > Linear.
Press Sweep Setup > Points > 201.
Press Sweep Setup > Sweep Type > Log Freq.
Press Sweep Setup > Osc Level > Osc Unit > Current.
Press Sweep Setup > Osc Level > Current Level > 1 > k/m.
Press Start > Start > 1 > G/n.
Press Stop > Stop > 3 > G/n.
Change the measurement conditions from the initial state of the E4991B:
Setup example for this measurement:
Parameter setting |
Setup example |
Initial state |
|
Measurement parameters |
Trace 1 |
Ls |
|Z| |
Trace 2 |
Q |
θz |
|
Display formats |
Trace 1 |
Linear |
Linear |
Trace 2 |
Linear |
Linear |
|
Measurement points |
201 points |
201 points |
|
Sweep parameter |
Oscillator level |
Frequency |
|
Sweep type |
Linear |
Linear |
|
Source mode |
Current |
Voltage |
|
CW frequency |
100 MHz |
1 MHz |
|
Sweep range (AC current) |
100 μA to 10 mA |
80 μA to 10.02 mA |
Preset the E4991B.
Press Display > Num of Traces > 2.
Press Meas > Trace 1 > Ls.
Press Format > Y-Axis > Lin.
Similarly, set Trace 2 > Q with Y-Axis > Linear.
Press Sweep Setup > Points > 201.
Press Sweep Setup > Sweep Type > Osc level > Osc Unit > Current.
Press Sweep Setup > Osc Level.
Press Sweep Setup > CW Freq > 100 > M/μ.
Press Start > Start > 100 > M/μ.
Press Stop > Stop > 10 > k/m.
When Option 001 is installed in the E4991B, DC bias can be applied to the DUT. First, change the measurement conditions from the initial state of the E4991B.
When measuring DC bias characteristics, you must set DC bias to ON after connecting the DUT to the test fixture.
Setup example for this measurement:
Parameter setting |
Setup example |
Initial state |
|
Measurement parameters |
Trace 1 |
Ls |
|Z| |
Trace 2 |
Q |
θz |
|
Display formats |
Trace 1 |
Linear |
Linear |
Trace 2 |
Linear |
Linear |
|
Measurement points |
201 points |
201 points |
|
Sweep parameter |
DC bias (current) |
Frequency |
|
Sweep type |
Linear |
Linear |
|
Source mode |
Current |
Voltage |
|
Oscillator level |
1 mA |
100 mV (2 mA) |
|
CW frequency |
100 MHz |
1 MHz |
|
Sweep range (DC current) |
100 μA to 50 mA |
0 A to 0 A |
Preset the E4991B.
Press Display > Num of Traces > 2.
Press Meas > Trace 1 > Ls.
Press Format > Y-Axis > Lin.
Similarly, set Trace 2 > Q with Y-Axis > Linear.
Press Sweep Setup > Points > 201.
Press Sweep Setup > DC Bias > DC Bias State > ON.
Press Sweep Setup > DC Bias > DC Bias Mode > Current.
Press Sweep Setup > Osc Level > Osc Unit > Current.
Press Sweep Setup > Osc Level > Current Level > 1 > k/m.
Press Sweep Setup > CW Freq > 100 > M/μ.
Press Start > Start > 100 > M/μ.
Press Stop > Stop > 50 > k/m.