Setting User Calibration Function

Loading Correction Data From CSV File

Load a correction data file written in CSV format to the E5052B. Follow these steps.

  1. Press System > Instrument Setup > Correction > Import Power Correction Table.

  2. Select a correction data file you want to load from the softkey file list.

    CSV files (extension:.csv) in the usercal folder in the F: drive are listed as softkeys.

  3. To load a correction data file from a floppy disk, other media, or another folder, follow these steps.

    1. Press System > Instrument Setup > Correction > File Dialog ....

    2. Select a correction data file you want to load from the file dialog box.

Turning On/Off User Correction Function

Reflect the loaded correction data. Follow these steps.

  1. Press System > Instrument Setup > Correction.

  2. Press Power Correction to turn on the user calibration function for correction data.