Bit Position |
Name |
Description |
0-3 |
Not used |
Always 0. |
4 |
Measurement |
Set to “1” during measurement. |
5-11 |
Not used |
Always 0. |
12 |
Operation user defined status register summary |
Set to “1” when one of the enable bits in operation user defined status register is set to “1.” |
13-15 |
Not used |
Always 0. |
Bit4 is set to “1” even in the “Waiting for Trigger” state when the trigger is set to “Ext/Video” in the transient measurement. This is because pre-triggering is performed in “Waiting for Trigger” state in the transient measurement.
Issuing the *CLS command will clear all bits from the operation status event register