Defines the settings for phase reference.
CALCulate:MEASure:Phase | PREFerence | XREFerence
|
Applicable Models: E505xA with opt S96301xB (Read-Write) Enable and disable the auto phase reference frequency. |
|
Parameters |
|
mnum |
Measurement number for each measurement. |
<Boolean> |
Choose from: OFF / 0: set the Phase Reference Frequency manually (Set frequency to) ON / 1: set the Phase Reference Frequency automatically (Measure Frequency at) |
Examples |
CALC:MEAS2:PHAS:PREF:AUTO OFF calculate:measure2:phase:preference:auto off |
Query Syntax |
CALCulate:MEASure<mnum>:PHASe:PREFerence:AUTO? |
Return Type |
Boolean |
OFF |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the "Set Frequency to" setting in the phase reference frequency. |
|
Parameters |
|
mnum |
Measurement number for each measurement. |
<value> |
Phase reference frequency |
Examples |
CALC:MEAS2:PHAS:PREF:FREQ 1E6 calculate:measure2:phase:preference:frequency |
Query Syntax |
CALCulate:MEASure<mnum>:PHASe:PREFerence:FREQuency? |
Return Type |
Numeric |
1E9 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the "Measure frequency at" setting in the phase reference frequency. |
|
Parameters |
|
mnum |
Measurement number for each measurement. |
<value> |
Phase reference in sec |
Examples |
CALC:MEAS2:PHAS:PREF:XPOS 0.0001 calculate:measure2:phase:preference:xposition 0.0001 |
Query Syntax |
CALCulate:MEASure<mnum>:PHASe:PREFerence:XPOSition? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the "Using Span of" setting in the phase reference frequency. |
|
Parameters |
|
mnum |
Measurement number for each measurement. |
<value> |
Phase span in sec |
Examples |
CALC:MEAS2:PHAS:PREF:XSP 0.001 calculate:measure2:phase:preference:xspan 0.001 |
Query Syntax |
CALCulate:MEASure<mnum>:PHASe:PREFerence:XSPan? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with opt S96301xB (Read-Write) Set and read the "Phase to Zero at" setting in the phase zero position. |
|
Parameters |
|
mnum |
Measurement number for each measurement. |
<value> |
Phase zero position in sec |
Examples |
CALC:MEAS2:PHAS:XREF:XPOS 0.001 calculate:measure2:phase:xreference:xposition 0.0001 |
Query Syntax |
CALCulate:MEASure<mnum>:PHASe:XREFerence:XPOSition? |
Return Type |
Numeric |
0 |
Applicable Models: E5055x with opt S96301xB (Read-Write) Set and read the "Using Span of" setting in the phase zero position. |
|
Parameters |
|
mnum |
Measurement number for each measurement. |
<value> |
Span in sec |
Examples |
CALC:MEAS2:PHAS:XREF:XSP 1e-4 calculate:measure2:phase:xreference:xspan 1e-4 |
Query Syntax |
CALCulate:MEASure<mnum>:PHASe:XREFerence:XSPan? |
Return Type |
Numeric |
0 |