Method (Write Only)
SCPI.IEEE4882.TRG
This command triggers the E5061B if the trigger source is set to GPIB/LAN (set to BUS with the SCPI.TRIGger.SEQuence.SOURce).
SCPI.TRIGger.SEQuence.SOURce = "bus"
SCPI.IEEE4882.TRG
No equivalent key is available on the front panel.
*TRG
10 OUTPUT 717;"*TRG"