The 2-port TRL calibration function lets you measure calibration data by connecting thru, reflection (open or short), line, or match calibration standards to (between) 2 desired test ports. This calibration provides the most accurate measurement for non-coaxial parts, using 12 error terms model in total for calibration in the same way as full 2-port calibration.
Before executing TRL calibration, you need to modify the calibration kit definition you use (or create a new one). For more information, see Changing the Calibration Kit Definition.
Press Channel Next/Channel Prev keys to select the channel for which you want to perform the calibration.
Press Cal key.
Click Calibrate > 2-Port TRL Cal.
Connect the port.(In the following procedure, test ports you select here are expressed as test ports x and y.)
Make a THRU or Line connection between ports x and y (between the connectors to which the DUT is to be connected).
Click Thru/Line > Port x-y Thru to execute a Thru/Line measurement.
Click Return.
Click Reflect > Port x Reflect to execute a reflection calibration for test port x.
Click Reflect > Port y Reflect to execute a reflection calibration for test port y.
Click Return.
Click Line/Match.
Click softkey to execute Line/Match calibration.
Click Return.
Click Done to complete TRL 2-port calibration. At this point, the calibration coefficient is calculated and saved. The error correction function is automatically turned ON.