Other topics about Data Output
As mentioned in Saving and Recalling Instrument State, you can save the instrument state of the E5063A into a file on the mass storage and then recall it later to reproduce that state.
There is no compatibility with other ENA series products.
The E5063A with option 011 (Time domain/Test wizard ) can recall the files saved by the E5063A without option 011.
The E5063A without option 011 (Time domain/Test wizard ) can recall the files saved by the E5063A with option 011. The settings for Transform /Gating are not used but they are kept.
|
Recalling by E5063A |
|||
245 |
285 |
2H5 |
||
Saved file by E5063A |
245 |
Y |
Y |
Y |
285 |
Y# |
Y |
Y |
|
2H5 |
Y# |
Y# |
Y |
Y: Recall is possible, Y# Warning message prompts "Incompatible recall file". Some state data may be lost.
Files that are saved with later firmware revisions may not be recalled by prior firmware revisions.
When files that are saved with different system spec. versions include calibration data, only states and trace data are recalled.
If you recall an incompatible file, an error occurs and the device recovers to the default setting.