Command
:SENSe<Ch>:CORRection:COEFficient:METHod:ERESponse <Value1>, <value2>
This command sets the calibration type to the enhanced response calibration between the two specified ports when the calibration coefficient data array is written with the :SENS:CORR:COEF command.
Parameter |
Value1 |
Description |
Port number for response port |
Data Type |
Numeric type (Integer) |
Range |
1 ~ 2 |
Parameter |
Value2 |
Description |
Port number for Stimulus port |
Data Type |
Numeric type (Integer) |
Range |
1 ~ 2 |
See Reading/Writing Error Coefficient.
No equivalent key is available on the front panel.
SCPI.SENSE(Ch).CORRECTION.COEFFICIENT.METHOD.ERESPONSE