Command
:SENSe<Ch>:CORRection:COEFficient:METHod[:RESPonse]:THRU <Value1>, <value2>
This command sets the calibration type to the response calibration (thru) between the two specified ports when the calibration coefficient data array is written with the :SENS:CORR:COEF command.
|
Parameter |
Value1 |
|
Description |
Port number for response port |
|
Data Type |
Numeric type (Integer) |
|
Range |
1 ~ 2 |
|
Parameter |
Value2 |
|
Description |
Port number for stimulus port |
|
Data Type |
Numeric type (Integer) |
|
Range |
1 ~ 2 |
See Reading/Writing Error Coefficient.
No equivalent key is available on the front panel.
SCPI.SENSE(Ch).CORRECTION.COEFFICIENT.METHOD.RESPONSE.THRU