:SENS<Ch>:CORR:COLL:CKIT

Type

Command/Query

Syntax

:SENSe<Ch>:CORRection:COLLect:CKIT[:SELect] <Value>

:SENSe<Ch>:CORRection:COLLect:CKIT[:SELect]?

Description

This command sets/gets the calibration kit.

Variable

Parameter

Value

Description

calibration kit number

Data Type

Numeric type (Integer)

Range

1 ~ 30

Preset Value

1

Query Response

{numeric}<newline><^END>

Examples

Dim Var as Long
Var= 1
Ena.WriteString ":SENS1:CORR:COLL:CKIT " & Str(Var), True
Ena.WriteString ":SENS1:CORR:COLL:CKIT?", True
Var=Ena.ReadNumber

Related Commands

:SENS:CORR:COLL:CKIT:ORD:LOAD

:SENS:CORR:COLL:CKIT:ORD:OPEN

:SENS:CORR:COLL:CKIT:ORD:SHOR

:SENS:CORR:COLL:CKIT:ORD:THRU

:SENS:CORR:COLL:CKIT:ORD:TRLL

:SENS:CORR:COLL:CKIT:ORD:TRLR

:SENS:CORR:COLL:CKIT:ORD:TRLT

Equivalent Softkey

No equivalent key is available on the front panel.

COM Command Syntax (Internal Control Only)

SCPI.SENSE(Ch).CORRECTION.COLLECT.CKIT.SELECT