Softkey |
Function |
2-1 (S21) |
Selects test port 2 (input) and test port 1 (output); Corresponds to the determination of S21 |
3-1 (S31) |
Selects test port 3 (input) and test port 1 (output); Corresponds to the determination of S31 |
4-1 (S41) |
Selects test port 4 (input) and test port 1 (output); Corresponds to the determination of S41 |
1-2 (S12) |
Selects test port 1 (input) and test port 2 (output); Corresponds to the determination of S12 |
3-2 (S32) |
Selects test port 3 (input) and test port 2 (output); Corresponds to the determination of S32 |
4-2 (S42) |
Selects test port 4 (input) and test port 2 (output); Corresponds to the determination of S42 |
1-3 (S13) |
Selects test port 1 (input) and test port 3 (output); Corresponds to the determination of S13 |
2-3 (S23) |
Selects test port 2 (input) and test port 3 (output); Corresponds to the determination of S23 |
4-3 (S43) |
Selects test port 4 (input) and test port 3 (output); Corresponds to the determination of S43 |
1-4 (S14) |
Selects test port 1 (input) and test port 4 (output); Corresponds to the determination of S14 |
2-4 (S24) |
Selects test port 2 (input) and test port 4 (output); Corresponds to the determination of S24 |
3-4 (S34) |
Selects test port 3 (input) and test port 4 (output); Corresponds to the determination of S34 |
An asterisk (*) in the upper right of the softkey indicates that the existing calibration coefficient will be cleared if you select the test port and execute acquisition of the calibration coefficient (clicking Done).