1-Port Calibration (reflection test)

Other topics about Basic Calibration

Overview

In 1-port calibration, calibration data are measured by connecting an OPEN standard, a SHORT standard, and a LOAD standard to the desired test port. This calibration effectively eliminates the frequency response reflection tracking error, directivity error, and source match error from the test setup in a reflection test using that port.

1-Port error model (1-port calibration)

Procedure

  1. Press Channel Next/Channel Prev keys to  select the channel for which you want to perform the calibration.

  2. Press Cal key.

  3. Click Calibrate > 1-Port Cal > Select Port.

  4. Select a test port (and corresponding S parameter) on which 1-port calibration will be performed.

  5. Connect an OPEN calibration standard to the selected test port (connector to which the DUT is to be connected).

  6. Click Open to start the calibration measurement.

  7. Connect a SHORT calibration standard to the selected test port (connector to which the DUT is to be connected).

  8. Click Short to start the calibration measurement.

  9. Connect a LOAD calibration standard to the selected test port (connector to which the DUT is to be connected).

  10. Click Load to start the calibration measurement.

  11. Click Done to terminate the 1-port calibration process. Upon pressing this key, calibration coefficients will be calculated and saved. The error correction function will also be automatically enabled.

Connecting the standard for 1-port calibration