Method (Write-only)
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.RChannel.ACQuire
This command executes receiver calibration (R channel only), for measurement the selected port (Pt) of selected channel (Ch).
Since information of power calibration for both the measurement port and the source port is used for error coefficient calculation, the precision of receiver calibration is improved by executing power calibration for both ports before executing receiver calibration.
When Calibrate R is selected, the Select Port setting is used and the Source Port setting is ignored.
SCPI.SENSe(1).CORRection.RECeiver(2).COLLect.RCHannel.ACQuire
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).STATe
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.ACQuire
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.TCHannel.ACQuire
Cal > Receiver Calibration > Calibrate R
:SENSe{[1]-160}:CORRection:RECeiver{[1]|2|3|4}:COLLect:RCHannel:ACQuire
10 OUTPUT 717;":SENS1:CORR:REC1:COLL:RCH:ACQ"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A