Calibration |
|
|
8753ES |
E5072A |
Calibration kit |
|
|
Available calibration kits:
|
See Calibration Kit Up to 11 user defined calibration kits can be registered. |
Calibration type |
|
|
|
|
Omission of the isolation measurement |
|
|
Possible |
Possible |
Power meter calibration |
|
|
Available |
Available |
Receiver calibration |
|
|
Available |
Available |
Adapter removal |
|
|
Available |
Not available |
Velocity factor |
|
|
Definable |
Definable |
Characteristic impedance of the measurement system |
|
|
Definable |
Definable with fixture simulator |
Port extension |
|
|
Definable for ports 1 and 2 and inputs A and B |
Definable for each port per channel |
Selection between chop sweep and alternate sweep |
|
|
Can be changed over. |
Chop sweep for traces in a channel and alternate sweep between channels |
Error correction |
|
|
On/Off setting is definable. |
Same |
Interpolating the calibration coefficient |
|
|
On/Off setting is definable. |
Always On. |
Defining the calibration kit |
Calibration class |
|
|
|
|
Class indication label |
|
Editable |
Editable |
|
Number of standards that can be registered in calibration classes |
|
Up to seven types |
Up to eight types |
|
Type of standard |
|
OPEN, SHORT, LOAD, and THRU |
OPEN, SHORT, LOAD, and THRU |
|
Standard parameters |
Common to standards |
|
Same
|
|
|
Unique to OPEN |
|
Same |
|
|
Unique to SHORT |
|
L0, L1, L2, L3 |
|
|
Unique to LOAD |
Type of standard (fixed, sliding, offset) |
Not available (the type of standard is treated as fixed load.) |
|
|
Unique to THRU |
Not available |
Unknown Thru |
|
|
Arbitrary Impedance |
|
Arbitrary Impedance |
|
Standard label |
|
Editable |
Same |
ECal |
Type of calibration |
|
|
|
|
Omission of isolation |
|
Allowed |
Allowed |
|
Manual THRU measurement |
|
Allowed |
Allowed |
|
Reading the frequency array |
|
Allowed |
Not allowed |
|
Reading the module information |
|
Allowed |
Not allowed |