Calibration

 

 

8753ES

E5072A 

Calibration kit

 

 

Available calibration kits:

  • 7 mm: 85031B

  • 3.5 mm: 85033C/D/E

  • N type (50 ohm): 85032B/E/F

  • N type (75 ohm): 85036B/E

  • 2.4 mm: 85056A/D

  • TRL 3.5 mm: 85052C

    User defined calibration kit

See Calibration  Kit

Up to 11 user defined calibration kits can be registered.

 

Calibration type

 

 

  • Response calibration

  • Response and isolation calibration

  • Enhanced response calibration

  • S11 1-port calibration

  • S22 1-port calibration

  • Full 2-port calibration

  • TRL/LRM Calibration

  • Receiver calibration

  • Power meter calibration

 

  • Response calibration (OPEN,SHORT, THRU)

  • Response and isolation calibration

    The response calibration includes isolation measurement as an option.

  • Enhanced response calibration

  • Full 1-port calibration

  • Full 2-port calibration

  • TRL/LRM Calibration

  • Power calibration

  • Receiver calibration

  • Vector-Mixer calibration

  • Scalar-Mixer calibration

  • Partial overwrite calibration

  • Unknown thru calibration

  • Adapter Removal/Insertion calibration

Omission of the isolation measurement

 

 

Possible
(Can be omitted by designating it using the softkeys on the front panel or sending a GPIB command)

Possible
(Isolation measurement is performed as an option accompanying a response calibration and 2-port calibration.)

Power meter calibration

 

 

Available

Available

Receiver calibration

 

 

Available

Available

Adapter removal

 

 

Available

Not available

Velocity factor

 

 

Definable

Definable

Characteristic impedance of the measurement system

 

 

Definable

Definable with fixture simulator

Port extension

 

 

Definable for ports 1 and 2 and inputs A and B

Definable for each port per channel

Selection between chop sweep and alternate sweep

 

 

Can be changed over.

Chop sweep for traces in a channel and alternate sweep between channels

Error correction

 

 

On/Off setting is definable.

Same

Interpolating the calibration coefficient

 

 

On/Off setting is definable.

Always On.

Defining the calibration kit

Calibration class

 

  • S11A (OPEN)

  • S11B (SHORT)

  • S11C (LOAD)

  • S22A (OPEN)

  • S22B (SHORT)

  • S22C (LOAD)

  • Forward Transmission

  • Forward Match

  • Reverse Transmission

  • Reverse Match

  • Response

  • Response & Isolation

  • TRL Thru

  • TRL Reflect

  • TRL Line/Match

  • OPEN

  • SHORT

  • LOAD

  • THRU
    (OPEN, SHORT, and LOAD are set for each port; THRU is set between ports)










     

  • TRL Thru

  • TRL Reflect

  • TRL Line/Match

 

Class indication label

 

Editable

Editable

 

Number of standards that can be registered in calibration classes

 

Up to seven types

Up to eight types

 

Type of standard 

 

OPEN, SHORT, LOAD, and THRU
Arbitrary Impedance

OPEN, SHORT, LOAD, and THRU
Arbitrary Impedance, None

 

Standard parameters

Common to standards

  • Offset delay

  • Offset loss

  • Offset characteristic impedance

  • Frequency range

  • Offset type (coaxial, waveguide)

Same 

 

 

 

 

Unique to OPEN

  • C0, C1, C2, C3

Same

 

 

Unique to SHORT

  • Not available

L0, L1, L2, L3

 

 

Unique to LOAD

Type of standard (fixed, sliding, offset)

Not available (the type of standard is treated as fixed load.)

 

 

Unique to THRU

Not available

Unknown Thru

 

 

Arbitrary Impedance

  • Type of standard (fixed, sliding)

  • Arbitrary Impedance

 Arbitrary Impedance

 

Standard label

 

Editable

Same

ECal

Type of calibration

 

  • Enhanced response calibration

  • S11 1-port calibration

  • S22 1-port calibration

  • Full 2-port calibration

  • Enhanced response calibration

  • Response (thru) calibration

  • Full 1-port calibration

  • Full 2-port calibration

 

Omission of isolation

 

Allowed

Allowed

 

Manual THRU measurement

 

Allowed

Allowed

 

Reading the frequency array

 

Allowed

Not allowed

 

Reading the module information

 

Allowed

Not allowed