Reset - is used to reset the soft front panel to its default settings.
Errors...
- is used to display reported errors.
To view the error log
Select Utilities > Errors.
Select Get Instrument Errors. All errors are displayed in the Errors dialog window.
To clear the error log
Close the Errors dialog window.
Re-open the Errors dialog window by selecting Utilities> Errors. This clears the errors and refreshes the Errors dialog window.
Self Test... - is used to run an instrument or module self test.
Select Utilities > Self Test.
All self test errors are displayed in the Self Test Results window.
If the unit passes its Self Test, "Self Test passed" is displayed.
After selft test, module is reset and the master power switch is turned off.
After selftest execution, open the C:\ProgramData\Keysight\M9379A\M9379A_{serial number}_SelfTestLog.txt
Module
temperature is shown at "MeasuredValue" in the line
of ParameterName "SwitchPath_Temp"
Code,IsInformational,Id,Slot,ParameterName,IsInclusiveLimits,LowerLimit,UpperLimit,MeasuredValue,MeasuredTemperature |
Turn on the master power switch again as necessary as it is turned off after the selftest.
Close the Self Test dialog window.
Re-open it by selecting Utilities > Self Test.
Checking the module temperature
Use Save As... to save the results of your Self Test as a text (.txt) file in a folder of your choice. If multiple files are saved in the same folder, the latest file will be at the bottom of the stack.
Firmware Update... - is used to update firmware on modules.
To Update the Firmware in a Module
Select a version of firmware for each module using the drop-down arrow.
Select Update to install available Firmware Updates.