Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to topic Go to topic Go to topic Go to topic Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description Go to description

7. Dual ARB Trigger

The image below shows a composite set of the Dual ARB Trigger parameters when connected to an M9381A/M9383A.

The image below shows a composite set of the Dual ARB Trigger parameters when connected to an M9383B/M9384B/M9484C.

Multi-Channel Synchronization Group Sync Source

(M9381A only)

Choice: PXI0 | PXI1 | PXI2 | PXI3 | PXI4 | PXI5 | PXI6 | PXI7

Default: PXI0

Specify the backplane bus that primary VSG used to send Group Synchronization Signal to trigger secondary VSGs. And the secondary VSGs wait on this bus line.

Multi-Channel Synchronization Secondary Sync Source

(M9381A only)

Choice: PXI0 | PXI1 | PXI2 | PXI3 | PXI4 | PXI5 | PXI6 | PXI7

Specify the backplane bus that secondary VSG used to send Secondary Synchronization Signal to trigger primary VSG for handshaking.

Trigger Type

Choice: Continuous | Single | Gated | Segment Advance

Choice (M9381A): Continuous | Single

Choice (VXT models): Continuous | Single | Segment Advance

Default: Continuous

This cell selects the triggering mode.

Trigger Type

(M9383B/M9384B/M9484C only)

Choice: Continuous | Single

Default: Continuous

Double-click or use the drop-down menu to select the triggering mode.

Continuous – selects the continuous triggering mode, which enables the signal generator to repeat the modulating signal indefinitely until you turn off the modulation format, change triggers, or select another waveform.

Single – sets up a waveform to play once after receiving a trigger.

Trigger Source

Choice: Trigger Key | External | Bus

Choice (M9381A): External | Bus

Choice: Trigger Key | External 2 | Bus

Default: External

Default (VXT models): External 2

This cell adjusts the trigger source applied to the PATT TRIG IN or AUX I/O connectors.

Trigger Source

(M9383B/M9384B/M9484C only)

Choice: Trigger Key | External | Bus | Immediate

Default: External

Adjusts the trigger source applied to the PATT TRIG IN or AUX I/O connectors.

Trigger Key − selects the front panel hardkey as the trigger source. After making this selection, press the Trigger hardkey to trigger a waveform.

External − enables you to trigger a waveform with an externally supplied trigger signal.

Immediate − enables immediate triggering of the waveform playing.

Bus − enables you to trigger a waveform by sending a command through the GPIB, LAN, or AUXILIARY IO (RS-232) input connector.

External Termination

(M9383B/M9384B/M9484C only)

Choice: 50 Ohm | High Impedance

Default: High Impedance

Select the way the trigger line is terminated in the hardware.

External Level

(M9383B/M9384B/M9484C only)

Range: -3.5 to 3.5

Default: 0.5

Specifies the threshold level in volts at which the trigger transitions states.

External Source

Choice: Pattern Trigger In 1 | Trigger Pattern In 2

Choice (M9381A): TRIG1 | PXI0 | PXI1 | PXI2 | PXI3 | PXI4 | PXI5 | PXI6 | PXI7

Default: Pattern Trigger In 1

This cell selects the rear-panel input for the external trigger signal. External must be selected as the trigger source.

For the M9381A PXIe VSG, double-click or use the drop-down menu to select the front-panel input for the external trigger signal. External must be selected as the trigger source.

External Polarity

Choice: Positive | Negative

Default: Negative

Default (VXT models): Positive

This toggles the polarity of the TTL signal at the rear panel TRIGGER OUT connector.

External Delay

Choice: On | Off

Default: Off

This cell turns the external trigger delay off or on. The selection is active only after selecting External as the trigger source.

External Delay Time

This parameter is labeled as External Trigger Delay for VXT models.

Range:

N5182A: 8 ns to 30 s, in 8 ns steps

All other models: 10 ns to 40 s, in 10 ns steps

Default:

8 ns or 10 ns

1.000000000 ms (VXT models)

This cell sets the amount of time to delay the signal generator's response to an external trigger, in 10 ns steps. The delay is a path (time) delay between the time the signal generator receives the trigger and when it responds to the trigger.

Continuous

Choice: Free Run | Trigger & Run | Reset & Run

Choice (M9381A): Free Run| Trigger & Run

Default: Free Run for 802.11a/b/g/j/p, 802.11n MxN, and 802.11n 1ARB with 2 SG system; Trigger & Run for 802.11n Mx2, Mx3, and Mx4 systems.

This cell selects the waveform's response to triggers when Continuous is the trigger mode (continuous waveform transmission.)

Single

Choice: Buffered Trigger | No Retrigger | Restart on Trigger

Default: Restart on Trigger for 802.11a/b/g/j/p, 802.11n MxN, and 802.11n 1ARB with 2 SG system; No Retrigger for 802.11n Mx2, Mx3, and Mx4 systems.

This cell configures the waveform's response to triggers while using the single trigger mode. The response selections determine whether the waveform accepts a trigger during playback, and if it does, how it responds to that trigger.

Segment Advance

Choice: Single | Continuous

Default: Continuous

This cell configures how play moves from segment to segment within a waveform sequence.

Gated

Choice: Active High | Active Low

Default: Active High

Sets the signal generator to the gated trigger mode, which causes the waveform (modulating signal) to repeatedly start and stop in response to an externally applied trigger signal. To use the gated trigger mode, you must also select External.

Sets the trigger active state for the gated trigger mode.

Trigger Pulse Generation

(M9381A only)

Choice: On | Off

Default:On

Double-click the cell or use the drop-down menu to select a trigger pulse generation state. This enables a trigger pulse generation after waveform downloading finished.

To enable and perform the function, Trigger Source should be set to External and Continuous should be set Trigger Run.

External Trigger Routing Bus

(M9381A only)

Choice: PXI0 | PXI1 | PXI2 | PXI3 | PXI4 | PXI5 | PXI6 | PXI7

Default: PXI6

External Hold Off Time

(M9383B/M9384B/M9484C only)

Range: 4 ns to 10 s

Default: 1 ms

Sets the amount of time in seconds to hold off until the next trigger.