Choice: Off | On
Default: On
Double-click or use the drop-down menu to enable or disable the automatic level control.
– The ALC circuit constantly monitors and controls the output power level.
– No automatic leveling occurs.
Choice: Auto | 100.00 Hz | 1.00 kHz | 10.00 kHz | 100.00 kHz
Default: Auto
Double-click or use the drop-down menu to select an ALC bandwidth.
This parameter is not available for all instrument model numbers and may be grayed out.
Choice: Fixed | Modulation
Default: Modulation
Double-click or use the drop-down menu to select a fixed or modulated reference signal when doing a power search.
– Power search uses a fixed 0.5 V reference.
– Power search uses the RMS value of the current I/Q modulation as the reference.
Set
to to enable this parameter.Choice: Normal | Minimum
Default: Normal
Double-click or use the drop-down menu to select the level of the output power control during a power search routine.
– Performs power searches without setting the internal step attenuator to its maximum setting. Glitches and power changes may be present during the search.
– Sets the signal generator's step attenuator to its maximum setting during power search to protect sensitive circuits.
Set
to to enable this parameter.