Last updated: 2010-06-08
A device-specific error indicates that the instrument has detected an error that occurred due to some improper operations, or possibly an abnormal hardware/firmware condition. For example, an attempt by the user to set an out-of-range value will generate a device specific error.
These are call processing errors generated by the core instrument.
Error Message |
Description |
+200 TD-SCDMA call processing error; Uplink DPCH midamble shift cannot be greater than midamble configuration |
This error message is displayed for information purposes only. |
+201 TD-SCDMA call processing error; Operation not allowed due to current call status |
If the Cell Operating Mode is "Active Cell", this setting can only be changed when the call status is Idle. |
+202 TD-SCDMA call processing error; Operation not allowed due to incorrect closed loop power control |
This action is only allowed when the closed loop power control mode is set to "Alternating bits". |
+203 TD-SCDMA call processing error; No response to Activate RB Test Mode Request |
During the "RB Test Mode Call Origination to CELL_DCH" procedure, the test set does not receive Activate Radio Bearer Test Mode Complete message from the UE within 2.5 seconds of sending the Activate Radio Bearer Test Mode message. |
+204 TD-SCDMA call processing error; No response to Close UE Test Loop |
During the "RB Test Mode Call Origination to CELL_DCH" procedure, the test set does not receive Close UE Test Loop Complete message from the UE within 2.5 seconds of sending the Close UE Test Loop message |
+205 TD-SCDMA call processing error; No response to RRC Connection Setup |
During the RRC Connection Setup procedure, the test set does not receive the RRC connection Complete message from the UE within 9 seconds of sending the RRC Connection Setup message |
+206 TD-SCDMA call processing error; No response to Radio Bearer Setup |
During the "RB Test Mode Call Origination to CELL_DCH" procedure, the test set does not receive Radio Bearer Setup Complete message from the UE within 9 seconds of sending the Radio Bearer Setup message |
+207 TD-SCDMA call processing error; No response to Identity Request |
During the identification procedure, the test set does not receive the Identity Response message from the UE within some seconds of sending the Identity Request message |
+208 TD-SCDMA call processing error; No response to Deactivate RB Test Mode |
This error message is displayed for information purposes only. |
+209 TD-SCDMA call disconnected; No response to RRC Connection Release |
During the Base Station Release from RB Test Mode in CELL_DCH procedure, the test set does not receive RRC Connection Release Complete message from the UE within 9 seconds of transmitting the RRC connection Release message |
+210 TD-SCDMA call disconnected; Call setup aborted |
This error message is displayed for information purposes only. |
+211 TD-SCDMA call processing error; No Physical Channel Reconfig Complete received |
During the Physical Channel Reconfiguration Procedure, the test set finds the uplink DPCH but fails to receive the appropriate "Reconfiguration Complete" message within 9 seconds of sending a physical channel reconfiguration message |
+212 TD-SCDMA call processing error; GPRS attach not completed |
During the GPRS Attach Procedure, the test sest fails to receive the GPRS Attach Complete message within 6 seconds of sending the GPRS Attach Accept message |
+213 TD-SCDMA call processing error; No response to Authentication Request |
During the authentication procedure, the test set fails to receive the Authentication (and Ciphering) Response or Authentication (and Ciphering) Failure message within certain seconds of sending the Authentication (and Ciphering) Request message |
+214 TD-SCDMA call processing error; No response to Security Mode Command |
During the authentication procedure, the test set fails to receive the Security Mode Complete message |
+215 TD-SCDMA call processing error; Authentication failed |
During the authentication procedure, the test set receives an Authentication (and Ciphering) Failure message |
+216 TD-SCDMA call processing error; Security Mode failed |
During the authentication procedure, the test set receives the Security Mode Failure message |
+217 TD-SCDMA call disconnected; No response to Open UE Test Loop |
During the Base Station Release from RB Test Mode in CELL_DCH procedure, the test set does not receive the Open UE Test Loop Complete message from the UE |
+218 TD-SCDMA call disconnected; Max no. RLC Retransmissions exceeded. Starting RLC Reset Proc. |
This error message is displayed for information purposes only. |
+219 TD-SCDMA call disconnected; No signal found on uplink DPCH |
During some call processing procedure (for example, the RRC Connection Setup or Physical Channel Reconfigure Procedure), the test set fails to find a valid DPCH |
+220 TD-SCDMA call disconnected; Signal lost on uplink DPCH |
When the Call Drop Timer State is "On", the test set finds a link failure has occurred due to loss of synchronization with the midamble in the Uplink DPCH |
+221 TD-SCDMA call processing error; Signal not found on uplink PRACH |
During the RRC Connection Setup procedure, the test set fails to receive a correct RACH message within 200 milliseconds after having received a valid UpPCH. |
+222 TD-SCDMA call processing error; UE Periodical Measurement Control Failure occurred |
The UE responds to the Measurement Control message with a Measurement Control Failure message. |
+224 TD-SCDMA call processing error; No response to CC Connect |
During the mobile initiated AMR service procedure, the test set does not receive the Connect Acknowledged message within the corresponding timer. |
+225 TD-SCDMA call disconnected; Bearer capability not supported |
|
+226 TD-SCDMA call processing error; No response to CC Release |
During the mobile station release from AMR voice procedure, the test set does not receive Release Complete message within the corresponding timer. |
+228 TD-SCDMA call processing error; No response to CC Disconnect |
During the test set release from AMR voice procedure, the test set does not receive the release message within the corresponding timer. |
+229 TD-SCDMA call processing error; No CC Call Confirmed received from the UE |
During the test set initiated AMR voice procedure, the test set does not receive the Call Confirmed message within the corresponding timer. |
+230 TD-SCDMA call processing error; No CC Alerting received from the UE |
During the test set initiated AMR voice procedure, the test set does not receive the Alerting message within the corresponding timer. |
+231 Call operation rejected; Call processing in progress |
The setting cannot be changed while in Active Cell Mode and the Call Status is not IDLE. |
+231 TD-SCDMA call processing error; No CC Connect received from the UE |
During the test set initiated AMR voice procedure, the test set does not receive the Connect message within the corresponding timer. |
+232 Call operation rejected; Protocol request ignored |
|
+233 TD-SCDMA call processing error; Call connection failed due to UE ’s capability or 8960’s configuration |
This error occurs when you are trying to set up an HSDPA call, but the Protocol Revision Control being set is not comply with the UE's capability. |
+234 TD-SCDMA call processing error; No response to Radio Bearer Release |
This error occurs when the test set does not receive Radio Beaer Release Complete message after sending the Radio Beaer Release message. |
+235 TD-SCDMA call processing error; Physical Channel Reconfig Failure received |
During the Physical Channel Reconfigure Procedure, the test set receives the Reconfiguration Failure message. |
+236 TD-SCDMA call processing error; Radio Bearer Setup Failure received |
During the RB Test Mode Call Origination to CELL_DCH procedure, the test set receives Radio Bearer Setup Fail message from the UE. |
+237 TD-SCDMA call processing error; No response to Location Update Accept |
During the Location Update Procedure, the test set does not receive the TMSI Reallocation Complete message from the UE within the required timer. |
+238 TD-SCDMA call disconnected; No response to RLC Poll. |
This is for information only. |
+239 TD-SCDMA call disconnected; RLC Reset Procedure Failed. |
This is for information only. |
+240 TD-SCDMA call processing error; Operation not allowed due to conflict H-RNTIs |
See H-RNTI Parameters. |
+241 TD-SCDMA call processing error; The simultaneous services feature is not supported |
The E1969A doesn’t support the simultaneous services
|