Description

These messages are intended to be displayed on the manual user interface only, and are not entered into the Error/Event Queue.

Error Message

Description

The function you requested is not yet available.

The test set does not have this capability.

IQ Calibration completed successfully for modulator <N>

<N> is the IQ modulator number that you are attempting to calibrate, <N> is 1 or 2.

IQ Calibration failed for modulator <N>

<N> is the IQ modulator number that you are attempting to calibrate, <N> is 1 or 2.

The function you requested is not available in this TA.

This function is used in another Test Application.

IQ Calibration for modulator 1 in progress (10 minutes). Call processing disabled.

This error is cleared by either the; IQ Calibration completed successfully for modulator <N>, or IQ Calibration failed for modulator <N>.

IQ Calibration for modulator 2 in progress (6 minutes). Call processing disabled.

This error is cleared by either the; IQ Calibration completed successfully for modulator <N>, or IQ Calibration failed for modulator <N>.

Instrument warning: Audio generator instrument has been closed.

The audio generator instrument was closed automatically by the test set.

Measurement warning: Audio Analyzer instrument has been closed.

The audio analyzer instrument was closed automatically by the test set.

Measurement warning: Analog audio measurement has been closed.

Analog audio measurements have been closed by the test set.

Protocol warning: Setting change has terminated the link with the DUT

Some setting change has caused the call to disconnect.

Warning; Call processing disabled

The call processing functions are not active because the test set is performing calibration operations.

IQ first modulator calibration X%

IQ modulator calibration in progress, X represents the percent complete.

IQ second modulator calibration X%

IQ modulator calibration in progress, X represents the percent complete.

WCDMA and FDD Manual User Messages

Error Message Description
WCDMA common measurements calibration completed successfully. The calibration procedure for WCDMA measurements completed without problems.
WCDMA common measurements calibration failed due to temperature drift. Wait for temperature to stabilize. The temperature in the test set was changing during calibration, causing the calibration to fail.
WCDMA common measurements calibration failed due to oven out of range. Indicates an internal problem during measurement calibration.

WCDMA common measurements calibration failed due to loopback switch problem.

Indicates an internal problem during measurement calibration.

WCDMA common measurements calibration failed due to under range condition. Indicates an internal problem during measurement calibration.
WCDMA common measurements calibration failed due to over range condition. Indicates an internal problem during measurement calibration.
WCDMA common measurements calibration failed due to correlation problem. Indicates an internal problem during measurement calibration.
WCDMA common measurements calibration failed. The measurement calibration procedure failed for an unspecified reason.
WCDMA common measurements calibration in progress. Do not apply RF power during calibration Tells you to not apply power during measurement calibration.
WCDMA common measurements calibration <x>%. Progress indicator for the measurement calibration routine.
FDD measurement warning; Thermal Power measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when thermal power measurements are initiated.
FDD measurement warning; Adjacent Channel Leakage Ratio measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when adjacent channel leakage ratio measurements are initiated.
FDD measurement warning; Waveform Quality measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when waveform quality measurements are initiated.
FDD measurement warning; Channel Power measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when channel power measurements are initiated.
FDD measurement warning; Loopback Bit Error Ratio measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when loopback bit error ratio measurements are initiated.
FDD measurement warning; Spectrum Emission Mask measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when spectrum emission mask measurements are initiated.
FDD measurement warning; Code Domain measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when code domain measurements are initiated.
FDD measurement warning; Occupied Bandwidth measurement has been closed. Indicates that a measurement has been inactivated because of a resource conflict. This message is typically generated when occupied bandwidth measurements are initiated.