Source Settings
Some of these settings are also available in the Instrument View. The Settings tables contain one row for each available output.
Basic Settings
Priority Mode
|
Sets the output priority mode (Voltage, Current, Power or Resistance). |
Settings
|
Sets the output Voltage, Current, Power or Resistance level. |
Limit
|
Limits the selected parameter at the specified value. |
Output Range
|
Sets the output range, for power modules with multiple ranges. |
Measurement Range
|
Sets the measurement range, for power modules with multiple ranges. |
Sensing and Coupling
Sensing
|
Sets the output sense to 2-wire or 4-wire. |
Coupling
|
Enables or disables the output coupling or synchronization between multiple output channels |
On Delay Time
|
Enter the On Delays for all outputs that will participate in the output on delay sequence. |
Off Delay Time
|
Enter the Off Delays for all outputs that will participate in the output off delay sequence.
|
OVP and OCP Settings
OVP
|
Enables Over Voltage Protection (OVP) with a corresponding voltage level. Depending on the PS model, you can also set a Delay Time. |
OCP
|
Enables Over Current Protection (OCP) with a corresponding level. Depending on the PS model, you can also set a Delay Time and specify if the start of the delay is initiated by any transition into CC mode, or only at the end of a settings change.
|
Clear
|
If a protection function is triggered, remove the condition that caused the protection fault, then click the output's Clear button to reset the protection. |