50 MHz Calibrator Amplitude Adjustment

This adjustment applies to the following models:

This adjustment adjusts the internal 50 MHz calibrator signal to be equal amplitude to an external 50 MHz, -25.00 dBm signal. The reference signal is provided by a calibrated, traceable oscillator.

Required Test Equipment

Test Equipment

Model Number1

50 MHz Calibrator

Z5602B Option H51

50 MHz Calibrator

  • For Option C35

Z5602B Option H35

50 MHz Calibrator

  • For Option 544

Z5602B Option H24

  1. Refer to the main Required Equipment table for a list of alternative models.

Connection Setups

50 MHz Calibrator Amplitude Adjustment Setup