N9048B PXE
This adjustment adjusts the internal 50 MHz calibrator signal to be equal amplitude to an external 50 MHz, -25.00 dBm signal. The reference signal is provided by a calibrated, traceable oscillator.
|
Test Equipment |
Model Number1 |
|---|---|
|
50 MHz Calibrator |
Z5602B Option H51 |
|
50 MHz Calibrator
|
Z5602B Option H35 |
|
50 MHz Calibrator
|
Z5602B Option H24 |
|