All adjustments will require the CFG Option. Refer to the CFG Option section for more information. |
During this adjustment, the analyzer locks to an Ext Ref 10 MHz signal via DAC. It stores DAC setting to use when not in External Reference mode.
This adjustment assumes that the 10 MHz reference is connected to the UUT. If the 10 MHz reference is not detected by the software, then the "standard" connection prompt is displayed. It is only necessary to connect the 10 MHz reference to complete the test. The other items in the "standard" connection prompt may be ignored.
|
Related test: |
Frequency Accuracy |
|
Test Equipment |
Recommended Model Number |
|---|---|
|
Freqency Standard |
Microsemi 5071A |
|
Cable, BNC |
8120-2582 |
|
|
The following notes apply to illustrations in the FieldFox Help:
|